Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Proc Inst Mech Eng H ; 224(2): 225-40, 2010.
Artigo em Inglês | MEDLINE | ID: mdl-20349816

RESUMO

A detailed review is given of the application of high-frequency ultrasound (HFUS) at frequencies of 20 MHz and above for high-resolution, cross-sectional imaging of biological soft tissue. The state of the art of HFUS imaging systems is discussed with respect to the underlying engineering concepts, system designs, and available transducer technology. Furthermore, the dependency of the spatial resolution on the system's parameters is analysed. Skin imaging, eye imaging, small animal imaging for preclinical research, and intravascular ultrasound in coronary arteries for arteriosclerotic disease diagnostics are presented as examples for the application of HFUS imaging in medical diagnostics. It is shown that, in the frame of the indicated applications, ultrasound in the frequency range 20-100MHz gives a good compromise between the contrary demands for a good spatial resolution and a sufficiently large penetration depth of ultrasound waves into the tissue. Scanning schemes for the imaging of tissue morphology are considered, including spatial compounding as a multidirectional imaging technique.


Assuntos
Angiografia/métodos , Aumento da Imagem/métodos , Microscopia Acústica/métodos , Oftalmoscopia/métodos , Angiografia/instrumentação , Desenho de Equipamento , Aumento da Imagem/instrumentação , Microscopia Acústica/instrumentação , Oftalmoscópios
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...