RESUMO
We report a case of pulmonary vein (PV) occlusion in a patient with a history of surgical atrial fibrillation ablation and left atrial appendage occlusion with unsuccessful endovascular management. Delayed diagnosis of PV stenosis post-ablation can make interventional treatment options less likely to be successful. (Level of Difficulty: Intermediate.).
RESUMO
BACKGROUND: Inside-out abrasion with externalization of sensing ring or high-voltage cables in St Jude Medical Riata implantable cardioverter-defibrillator leads has been reported. The prevalence of extruded cables, rate of electrical abnormalities, and predictors of failure in Riata leads are unknown. OBJECTIVES: To estimate the incidence of lead failure in the St Jude Medical Riata implantable cardioverter-defibrillator leads and to propose a standard for the fluoroscopic assessment of insulation breakdown. METHODS: Patients undergoing cine-fluoroscopy on Riata implantable cardioverter-defibrillator leads at our institution before January 25, 2012, were included (n = 87). Leads were graded as types 0-3 (0 = normal, 1 = abnormal conductor spacing, 2 ≤1 cm cable extrusion, 3 = >1 cm length extrusion). Comparison to extracted leads (n = 15) was documented. Device interrogation data were used for electrical analysis. RESULTS: The mean time from implant was 5.9 ± 3.45 years. Structural lead failure with externalized cables was seen in 33.3% (29 of 87) of the patients. Thirty-one percent (9 of 29) of the leads with exposed cables showed electrical failure, and 29.7% (19 of 64) of the leads with normal electrical data contained externalized cables. Time from implant ≥5 years predicted structural lead failure (P < 0.05). X-ray grade compared with extracted leads demonstrated a sensitivity and specificity of 86% and 100%, respectively. CONCLUSIONS: Cine-fluoroscopy using a simple scale correlated with the structural integrity of extracted Riata leads. A high percentage of leads with extrusion showed electrical failure. Leads ≥5 years from implant showed a high rate of externalized cables. A large independent multicenter study to determine the prevalence and clinical sequelae of Riata lead failures is warranted.