1.
J Bone Joint Surg Br ; 89(3): 291-5, 2007 Mar.
Artigo em Inglês | MEDLINE | ID: mdl-17356136

RESUMO

We reviewed the initial post-operative radiographs of the Trident acetabulum and identified a problem with seating of the metal-backed ceramic liner. We identified 117 hips in 113 patients who had undergone primary total hip replacement using the Trident shell with a metal-backed alumina liner. Of these, 19 (16.4%) were noted to have incomplete seating of the liner, as judged by plain anteroposterior and lateral radiographs. One case of complete liner dissociation necessitating early revision was not included in the prevalence figures. One mis-seated liner was revised in the early post-operative period and two that were initially incompletely seated were found on follow-up radiographs to have become correctly seated. There may be technical issues with regard to the implanting of this prosthesis of which surgeons should be aware. However, there is the distinct possibility that the Trident shell deforms upon implantation, thereby preventing complete seating of the liner.


Assuntos
Acetábulo/diagnóstico por imagem , Artroplastia de Quadril/instrumentação , Articulação do Quadril/diagnóstico por imagem , Prótese de Quadril , Acetábulo/cirurgia , Adulto , Idoso , Artroplastia de Quadril/métodos , Estudos de Coortes , Feminino , Articulação do Quadril/cirurgia , Humanos , Artropatias/cirurgia , Masculino , Pessoa de Meia-Idade , Cuidados Pós-Operatórios/métodos , Desenho de Prótese , Falha de Prótese , Radiografia , Reoperação
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...