1.
Artif Organs ; 39(8): 681-90, 2015 Aug.
Artigo em Inglês | MEDLINE | ID: mdl-26146861

RESUMO

This study in five large greyhound dogs implanted with a VentrAssist left ventricular assist device focused on identification of the precise site and physiological changes induced by or underlying the complication of left ventricular suction. Pressure sensors were placed in left and right atria, proximal and distal left ventricle, and proximal aorta while dual perivascular and tubing ultrasonic flow meters measured blood flow in the aortic root and pump outlet cannula. When suction occurred, end-systolic pressure gradients between proximal and distal regions of the left ventricle on the order of 40-160 mm Hg indicated an occlusive process of variable intensity in the distal ventricle. A variable negative flow difference between end systole and end diastole (0.5-3.4 L/min) was observed. This was presumably mediated by variable apposition of the free and septal walls of the ventricle at the pump inlet cannula orifice which lasted approximately 100 ms. This apposition, by inducing an end-systolic flow deficit, terminated the suction process by relieving the imbalance between pump requirement and delivery from the right ventricle. Immediately preceding this event, however, unnaturally low end-systolic pressures occurred in the left atrium and proximal left ventricle which in four dogs lasted for 80-120 ms. In one dog, however, this collapse progressed to a new level and remained at approximately -5 mm Hg across four heart beats at which point suction was relieved by manual reduction in pump speed. Because these pressures were associated with a pulmonary capillary wedge pressure of -5 mm Hg as well, they indicate total collapse of the entire pulmonary venous system, left atrium, and left ventricle which persisted until pump flow requirement was relieved by reducing pump speed. We suggest that this collapse caused the whole vascular region from pulmonary capillaries to distal left ventricle to behave as a Starling resistance which further reduced right ventricular output thus contributing to a major reduction in pump flow. We contend that similar complications of manual speed control also occur in the human subject and remain a major unsolved problem in the clinical management of patients implanted with rotary blood pumps.


Assuntos
Coração Auxiliar/efeitos adversos , Hemodinâmica , Falha de Prótese , Disfunção Ventricular Esquerda/etiologia , Função Ventricular Esquerda , Animais , Modelos Animais de Doenças , Cães , Modelos Cardiovasculares , Desenho de Prótese , Volume Sistólico , Fatores de Tempo , Transdutores de Pressão , Resistência Vascular , Disfunção Ventricular Esquerda/diagnóstico , Disfunção Ventricular Esquerda/fisiopatologia , Pressão Ventricular
2.
Artif Organs ; 39(2): E24-35, 2015 Feb.
Artigo em Inglês | MEDLINE | ID: mdl-25345482

RESUMO

The present study investigates the response of implantable rotary blood pump (IRBP)-assisted patients to exercise and head-up tilt (HUT), as well as the effect of alterations in the model parameter values on this response, using validated numerical models. Furthermore, we comparatively evaluate the performance of a number of previously proposed physiologically responsive controllers, including constant speed, constant flow pulsatility index (PI), constant average pressure difference between the aorta and the left atrium, constant average differential pump pressure, constant ratio between mean pump flow and pump flow pulsatility (ratioP I or linear Starling-like control), as well as constant left atrial pressure ( P l a ¯ ) control, with regard to their ability to increase cardiac output during exercise while maintaining circulatory stability upon HUT. Although native cardiac output increases automatically during exercise, increasing pump speed was able to further improve total cardiac output and reduce elevated filling pressures. At the same time, reduced venous return associated with upright posture was not shown to induce left ventricular (LV) suction. Although P l a ¯ control outperformed other control modes in its ability to increase cardiac output during exercise, it caused a fall in the mean arterial pressure upon HUT, which may cause postural hypotension or patient discomfort. To the contrary, maintaining constant average pressure difference between the aorta and the left atrium demonstrated superior performance in both exercise and HUT scenarios. Due to their strong dependence on the pump operating point, PI and ratioPI control performed poorly during exercise and HUT. Our simulation results also highlighted the importance of the baroreflex mechanism in determining the response of the IRBP-assisted patients to exercise and postural changes, where desensitized reflex response attenuated the percentage increase in cardiac output during exercise and substantially reduced the arterial pressure upon HUT.


Assuntos
Simulação por Computador , Exercício Físico , Coração Auxiliar , Hemodinâmica , Modelos Cardiovasculares , Pressão Sanguínea , Débito Cardíaco , Humanos
3.
Artif Organs ; 36(9): 787-96, 2012 Sep.
Artigo em Inglês | MEDLINE | ID: mdl-22626056

RESUMO

A clinically intuitive physiologic controller is desired to improve the interaction between implantable rotary blood pumps and the cardiovascular system. This controller should restore the Starling mechanism of the heart, thus preventing overpumping and underpumping scenarios plaguing their implementation. A linear Starling-like controller for pump flow which emulated the response of the natural left ventricle (LV) to changes in preload was then derived using pump flow pulsatility as the feedback variable. The controller could also adapt the control line gradient to accommodate longer-term changes in cardiovascular parameters, most importantly LV contractility which caused flow pulsatility to move outside predefined limits. To justify the choice of flow pulsatility, four different pulsatility measures (pump flow, speed, current, and pump head pressure) were investigated as possible surrogates for LV stroke work. Simulations using a validated numerical model were used to examine the relationships between LV stroke work and these measures. All were approximately linear (r(2) (mean ± SD) = 0.989 ± 0.013, n = 30) between the limits of ventricular suction and opening of the aortic valve. After aortic valve opening, the four measures differed greatly in sensitivity to further increases in LV stroke work. Pump flow pulsatility showed more correspondence with changes in LV stroke work before and after opening of the aortic valve and was least affected by changes in the LV and right ventricular (RV) contractility, blood volume, peripheral vascular resistance, and heart rate. The system (flow pulsatility) response to primary changes in pump flow was then demonstrated to be appropriate for stable control of the circulation. As medical practitioners have an instinctive understanding of the Starling curve, which is central to the synchronization of LV and RV outputs, the intuitiveness of the proposed Starling-like controller will promote acceptance and enable rational integration into patterns of hemodynamic management.


Assuntos
Coração Auxiliar , Fluxo Pulsátil , Função Ventricular Esquerda , Humanos , Modelos Cardiovasculares