Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 2 de 2
Filtrar
Mais filtros










Base de dados
Assunto principal
Intervalo de ano de publicação
1.
Rev Sci Instrum ; 93(8): 083701, 2022 Aug 01.
Artigo em Inglês | MEDLINE | ID: mdl-36050103

RESUMO

We describe our miniature laser powder bed fusion (L-PBF) system for in situ synchrotron x-ray micro-computed tomography (XCT) at the European Synchrotron Radiation Facility. This replicator was designed to extend the characterization of L-PBF to 3D. This instrument fills in a technical gap because the existing replicators were mostly designed to shed light on the dynamic mechanisms involved in molten pool formation but, therefore, suffered from a lack of 3D information. Technical details regarding the setup and beamline integration are given. Experimental validations via post-mortem XCT scans and in situ scans acquired during experiments conducted at the BM05 beamline of the European Synchrotron Radiation Facility are provided. Based on a few illustrative examples, we show that such a replicator opens the path to collect key 3D information that to date could not be available. Our miniature instrument complements the other replicators developed in the world by other research groups that enable operando x-ray imaging (radiography) and operando x-ray diffraction.


Assuntos
Síncrotrons , Pós , Difração de Raios X , Microtomografia por Raio-X , Raios X
2.
J Synchrotron Radiat ; 28(Pt 2): 530-537, 2021 Mar 01.
Artigo em Inglês | MEDLINE | ID: mdl-33650566

RESUMO

Metallic materials processing such as rolling, extrusion or forging often involves high-temperature deformation. Usually under such conditions the samples are characterized post mortem, under pseudo in situ conditions with interrupted tests, or in situ with a limited strain rate. A full in situ 3D characterization, directly during high-temperature deformation with a prescribed strain-rate scheme, requires a dedicated sample environment and a dedicated image-analysis workflow. A specific sample environment has been developed to enable highly controlled (temperature and strain rate) high-temperature deformation mechanical testing to be conducted while performing in situ tomography on a synchrotron beamline. A dedicated digital volume correlation algorithm is used to estimate the strain field and track pores while the material endures large deformations. The algorithm is particularly suitable for materials with few internal features when the deformation steps between two images are large. An example of an application is provided: a high-temperature compression test on a porous aluminium alloy with individual pore tracking with a specific strain-rate scheme representative of rolling conditions.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...