J Instrum ; 8(2): C02033, 2013 Feb.
Artigo em Inglês | MEDLINE | ID: mdl-25540669

RESUMO

We have designed and built a large-area 1cm × 1cm position-sensitive solid-state photomultiplier (PS-SSPM) for use in detector design for medical imaging applications. Our new large-area PS-SSPM concept implements resistive network between the micro-pixels, which are photodiodes operated in Geiger mode, called Geiger Photodiodes (GPDs), to provide continuous position sensitivity. Here we present imaging and timing performance of the large-area PS-SSPM for different temperatures and operating biases to find the optimum operating parameters for the device in imaging applications. A detector module was built by coupling a polished 8×8 LYSO array, with 1×1×20 mm3 elements, to a 1×1 cm2 PS-SSPM. Flood images recorded at room temperature show good crystal separation as all 64 elements were separated from each other. Cooling the device at 10 °C showed significant improvement. The device optimum bias voltage was ~4.5V over breakdown voltage. The coincidence timing resolution was improved significantly by increasing the operating bias, as well as by lowering the temperature to 0 °C. Results show excellent imaging performance and good timing response with a large-area PS-SSPM device.