NMR Biomed ; : e5180, 2024 May 22.
Artigo em Inglês | MEDLINE | ID: mdl-38775032

RESUMO

Ultrahigh field magnetic resonance imaging (MRI) (≥ 7 T) has the potential to provide superior spatial resolution and unique image contrast. Apart from radiofrequency transmit inhomogeneities in the body at this field strength, imaging of the upper abdomen faces additional challenges associated with motion-induced ghosting artifacts. To address these challenges, the goal of this work was to develop a technique for high-resolution free-breathing upper abdominal MRI at 7 T with a large field of view. Free-breathing 3D gradient-recalled echo (GRE) water-excited radial stack-of-stars data were acquired in seven healthy volunteers (five males/two females, body mass index: 19.6-24.8 kg/m2) at 7 T using an eight-channel transceive array coil. Two volunteers were also examined at 3 T. In each volunteer, the liver and kidney regions were scanned in two separate acquisitions. To homogenize signal excitation, the time-interleaved acquisition of modes (TIAMO) method was used with personalized pairs of B1 shims, based on a 23-s Cartesian fast low angle shot (FLASH) acquisition. Utilizing free-induction decay navigator signals, respiratory-gated images were reconstructed at a spatial resolution of 0.8 × 0.8 × 1.0 mm3. Two experienced radiologists rated the image quality and the impact of B1 inhomogeneity and motion-related artifacts on multipoint scales. The images of all volunteers showcased effective water excitation and were accurately corrected for respiratory motion. The impact of B1 inhomogeneity on image quality was minimal, underscoring the efficacy of the multitransmit TIAMO shim. The high spatial resolution allowed excellent depiction of small structures such as the adrenal glands, the proximal ureter, the diaphragm, and small blood vessels, although some streaking artifacts persisted in liver image data. In direct comparisons with 3 T performed for two volunteers, 7-T acquisitions demonstrated increases in signal-to-noise ratio of 77% and 58%. Overall, this work demonstrates the feasibility of free-breathing MRI in the upper abdomen at submillimeter spatial resolution at a magnetic field strength of 7 T.

2.
Magn Reson Med ; 2024 May 10.
Artigo em Inglês | MEDLINE | ID: mdl-38733068

RESUMO

PURPOSE: To address the limitations of spinal cord imaging at ultra-high field (UHF) due to time-consuming parallel transmit (pTx) adjustments. This study introduces calibration-free offline computed universal shim modes that can be applied seamlessly for different pTx RF coils and spinal cord target regions, substantially enhancing spinal cord imaging efficiency at UHF. METHODS: A library of channel-wise relative B 1 + $$ {B}_1^{+} $$ maps for the cervical spinal cord (six datasets) and thoracic and lumbar spinal cord (nine datasets) was constructed to optimize transmit homogeneity and efficiency for these regions. A tailored B0 shim was optimized for the cervical spine to enhance spatial magnetic field homogeneity further. The performance of the universal shims was validated using absolute saturation based B 1 + $$ {B}_1^{+} $$ mapping and high-resolution 2D and 3D multi-echo gradient-recalled echo (GRE) data to assess the image quality. RESULTS: The proposed universal shims demonstrated a 50% improvement in B 1 + $$ {B}_1^{+} $$ efficiency compared to the default (zero phase) shim mode. B 1 + $$ {B}_1^{+} $$ homogeneity was also improved by 20%. The optimized universal shims achieved performance comparable to subject-specific pTx adjustments, while eliminating the need for lengthy pTx calibration times, saving about 10 min per experiment. CONCLUSION: The development of universal shims represents a significant advance by eliminating time-consuming subject-specific pTx adjustments. This approach is expected to make UHF spinal cord imaging more accessible and user-friendly, particularly for non-pTx experts.

3.
Magn Reson Med ; 2024 Mar 05.
Artigo em Inglês | MEDLINE | ID: mdl-38440957

RESUMO

PURPOSE: To research and evaluate the performance of broadband tailored kT-point pulses (TP) and universal pulses (UP) for homogeneous excitation of the human heart at 7T. METHODS: Relative 3D B 1 + $$ {\mathrm{B}}_1^{+} $$ -maps of the thorax were acquired from 29 healthy volunteers. TP and UP were designed using the small-tip-angle approximation for a different composition of up to seven resonance frequencies. TP were computed for each of the 29 B 1 + $$ {\mathrm{B}}_1^{+} $$ -maps, and UPs were calculated using 22 B 1 + $$ {\mathrm{B}}_1^{+} $$ -maps and tested in seven testcases. The performance of the pulses was analyzed using the coefficient of variation (CV) in the 3D heart volumes. The 3D gradient-echo (GRE) scans were acquired for the seven testcases to qualitatively validate the B 1 + $$ {\mathrm{B}}_1^{+} $$ -predictions. RESULTS: Single- and double-frequency optimized pulses achieved homogeneity in flip angle (FA) for the frequencies they were optimized for, while the broadband pulses achieved uniformity in FA across a 1300 Hz frequency range. CONCLUSION: Broadband TP and UP can be used for homogeneous excitation of the heart volume across a 1300 Hz frequency range, including the water and the main six fat peaks, or with longer pulse durations and higher FAs for a smaller transmit bandwidth. Moreover, despite large inter-volunteer variations, broadband UP can be used for calibration-free 3D heart FA homogenization in time-critical situations.

4.
Magn Reson Med ; 91(5): 1994-2009, 2024 May.
Artigo em Inglês | MEDLINE | ID: mdl-38174601

RESUMO

PURPOSE: Traditional phase-contrast MRI is affected by displacement artifacts caused by non-synchronized spatial- and velocity-encoding time points. The resulting inaccurate velocity maps can affect the accuracy of derived hemodynamic parameters. This study proposes and characterizes a 3D radial phase-contrast UTE (PC-UTE) sequence to reduce displacement artifacts. Furthermore, it investigates the displacement of a standard Cartesian flow sequence by utilizing a displacement-free synchronized-single-point-imaging MR sequence (SYNC-SPI) that requires clinically prohibitively long acquisition times. METHODS: 3D flow data was acquired at 3T at three different constant flow rates and varying spatial resolutions in a stenotic aorta phantom using the proposed PC-UTE, a Cartesian flow sequence, and a SYNC-SPI sequence as reference. Expected displacement artifacts were calculated from gradient timing waveforms and compared to displacement values measured in the in vitro flow experiments. RESULTS: The PC-UTE sequence reduces displacement and intravoxel dephasing, leading to decreased geometric distortions and signal cancellations in magnitude images, and more spatially accurate velocity quantification compared to the Cartesian flow acquisitions; errors increase with velocity and higher spatial resolution. CONCLUSION: PC-UTE MRI can measure velocity vector fields with greater accuracy than Cartesian acquisitions (although pulsatile fields were not studied) and shorter scan times than SYNC-SPI. As such, this approach is superior to traditional Cartesian 3D and 4D flow MRI when spatial misrepresentations cannot be tolerated, for example, when computational fluid dynamics simulations are compared to or combined with in vitro or in vivo measurements, or regional parameters such as wall shear stress are of interest.


Assuntos
Estenose da Valva Aórtica , Imageamento por Ressonância Magnética , Humanos , Imageamento por Ressonância Magnética/métodos , Hemodinâmica , Imagens de Fantasmas , Artefatos , Velocidade do Fluxo Sanguíneo , Imageamento Tridimensional/métodos
5.
Magn Reson Med ; 91(4): 1659-1675, 2024 Apr.
Artigo em Inglês | MEDLINE | ID: mdl-38031517

RESUMO

PURPOSE: To investigate safety and performance aspects of parallel-transmit (pTx) RF control-modes for a body coil at B 0 ≤ 3 T $$ {B}_0\le 3\mathrm{T} $$ . METHODS: Electromagnetic simulations of 11 human voxel models in cardiac imaging position were conducted for B 0 = 0.5 T $$ {B}_0=0.5\mathrm{T} $$ , 1.5 T $$ 1.5\mathrm{T} $$ and 3 T $$ 3\mathrm{T} $$ and a body coil with a configurable number of transmit channels (1, 2, 4, 8, 16). Three safety modes were considered: the 'SAR-controlled mode' (SCM), where specific absorption rate (SAR) is limited directly, a 'phase agnostic SAR-controlled mode' (PASCM), where phase information is neglected, and a 'power-controlled mode' (PCM), where the voltage amplitude for each channel is limited. For either mode, safety limits were established based on a set of 'anchor' simulations and then evaluated in 'target' simulations on previously unseen models. The comparison allowed to derive safety factors accounting for varying patient anatomies. All control modes were compared in terms of the B 1 + $$ {B}_1^{+} $$ amplitude and homogeneity they permit under their respective safety requirements. RESULTS: Large safety factors (approximately five) are needed if only one or two anchor models are investigated but they shrink with increasing number of anchors. The achievable B 1 + $$ {B}_1^{+} $$ is highest for SCM but this advantage is reduced when the safety factor is included. PCM appears to be more robust against variations of subjects. PASCM performance is mostly in between SCM and PCM. Compared to standard circularly polarized (CP) excitation, pTx offers minor B 1 + $$ {B}_1^{+} $$ improvements if local SAR limits are always enforced. CONCLUSION: PTx body coils can safely be used at B 0 ≤ 3 T $$ {B}_0\le 3\mathrm{T} $$ . Uncertainties in patient anatomy must be accounted for, however, by simulating many models.


Assuntos
Coração , Imageamento por Ressonância Magnética , Humanos , Imageamento por Ressonância Magnética/métodos , Simulação por Computador , Coração/diagnóstico por imagem , Imagens de Fantasmas , Ondas de Rádio
6.
MAGMA ; 36(2): 191-210, 2023 Apr.
Artigo em Inglês | MEDLINE | ID: mdl-37029886

RESUMO

Multiple sites within Germany operate human MRI systems with magnetic fields either at 7 Tesla or 9.4 Tesla. In 2013, these sites formed a network to facilitate and harmonize the research being conducted at the different sites and make this technology available to a larger community of researchers and clinicians not only within Germany, but also worldwide. The German Ultrahigh Field Imaging (GUFI) network has defined a strategic goal to establish a 14 Tesla whole-body human MRI system as a national research resource in Germany as the next progression in magnetic field strength. This paper summarizes the history of this initiative, the current status, the motivation for pursuing MR imaging and spectroscopy at such a high magnetic field strength, and the technical and funding challenges involved. It focuses on the scientific and science policy process from the perspective in Germany, and is not intended to be a comprehensive systematic review of the benefits and technical challenges of higher field strengths.


Assuntos
Imageamento por Ressonância Magnética , Imagem Corporal Total , Humanos , Imageamento por Ressonância Magnética/métodos , Espectroscopia de Ressonância Magnética , Imagem Corporal Total/métodos , Alemanha , Campos Magnéticos
7.
MAGMA ; 36(2): 257-277, 2023 Apr.
Artigo em Inglês | MEDLINE | ID: mdl-36920549

RESUMO

OBJECTIVE: To examine the feasibility of human cardiac MR (CMR) at 14.0 T using high-density radiofrequency (RF) dipole transceiver arrays in conjunction with static and dynamic parallel transmission (pTx). MATERIALS AND METHODS: RF arrays comprised of self-grounded bow-tie (SGBT) antennas, bow-tie (BT) antennas, or fractionated dipole (FD) antennas were used in this simulation study. Static and dynamic pTx were applied to enhance transmission field (B1+) uniformity and efficiency in the heart of the human voxel model. B1+ distribution and maximum specific absorption rate averaged over 10 g tissue (SAR10g) were examined at 7.0 T and 14.0 T. RESULTS: At 14.0 T static pTx revealed a minimum B1+ROI efficiency of 0.91 µT/√kW (SGBT), 0.73 µT/√kW (BT), and 0.56 µT/√kW (FD) and maximum SAR10g of 4.24 W/kg, 1.45 W/kg, and 2.04 W/kg. Dynamic pTx with 8 kT points indicate a balance between B1+ROI homogeneity (coefficient of variation < 14%) and efficiency (minimum B1+ROI > 1.11 µT/√kW) at 14.0 T with a maximum SAR10g < 5.25 W/kg. DISCUSSION: MRI of the human heart at 14.0 T is feasible from an electrodynamic and theoretical standpoint, provided that multi-channel high-density antennas are arranged accordingly. These findings provide a technical foundation for further explorations into CMR at 14.0 T.


Assuntos
Coração , Imageamento por Ressonância Magnética , Humanos , Coração/diagnóstico por imagem , Simulação por Computador , Ondas de Rádio , Imagens de Fantasmas , Desenho de Equipamento
8.
NMR Biomed ; 36(7): e4900, 2023 07.
Artigo em Inglês | MEDLINE | ID: mdl-36624556

RESUMO

To protect implant carriers in MRI from excessive radiofrequency (RF) heating it has previously been suggested to assess that hazard via sensors on the implant. Other work recommended parallel transmission (pTx) to actively mitigate implant-related heating. Here, both ideas are integrated into one comprehensive safety concept where native pTx safety (without implant) is ensured by state-of-the-art field simulations and the implant-specific hazard is quantified in situ using physical sensors. The concept is demonstrated by electromagnetic simulations performed on a human voxel model with a simplified spinal-cord implant in an eight-channel pTx body coil at 3 T . To integrate implant and native safety, the sensor signal must be calibrated in terms of an established safety metric (e.g., specific absorption rate [SAR]). Virtual experiments show that E -field and implant-current sensors are well suited for this purpose, while temperature sensors require some caution, and B 1 probes are inadequate. Based on an implant sensor matrix Q s , constructed in situ from sensor readings, and precomputed native SAR limits, a vector space of safe RF excitations is determined where both global (native) and local (implant-related) safety requirements are satisfied. Within this safe-excitation subspace, the solution with the best image quality in terms of B 1 + magnitude and homogeneity is then found by a straightforward optimization algorithm. In the investigated example, the optimized pTx shim provides a 3-fold higher mean B 1 + magnitude compared with circularly polarized excitation for a maximum implant-related temperature increase ∆ T imp ≤ 1 K . To date, sensor-equipped implants interfaced to a pTx scanner exist as demonstrator items in research labs, but commercial devices are not yet within sight. This paper aims to demonstrate the significant benefits of such an approach and how this could impact implant-related RF safety in MRI. Today, the responsibility for safe implant scanning lies with the implant manufacturer and the MRI operator; within the sensor concept, the MRI manufacturer would assume much of the operator's current responsibility.


Assuntos
Temperatura Alta , Ondas de Rádio , Humanos , Simulação por Computador , Imagens de Fantasmas , Imageamento por Ressonância Magnética/métodos
9.
Phys Med ; 105: 102514, 2023 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-36608390

RESUMO

PURPOSE: Assess and optimise acquisition parameters for continuous cardiac Magnetic Resonance Fingerprinting (MRF). METHODS: Different acquisition schemes (flip angle amplitude, lobe size, T2-preparation pulses) for cardiac MRF were assessed in simulations and phantom and demonstrated in one healthy volunteer. Three different experimental designs were evaluated using central composite and fractional factorial designs. Relative errors for T1 and T2 were calculated for a wide range of realistic T1 and T2 value combinations. The effect of different designs on the accuracy of T1 and T2 was assessed using response surface modelling and Cohen's f calculations. RESULTS: Larger flip angle amplitudes lead to an improvement of T2 accuracy and precision for simulations and phantom experiments. Similar effects could also be shown qualitatively in in-vivo scans. Accuracy and precision of T1 were robust to different design parameters with improved values for faster flip angle variation. Cohen's f showed that T2-preparation pulses influence the accuracy of T2. The number of pulses used is the most important parameter. Without T2-preparation pulses, RMSE were 3.0 ± 8.09 % for T1 and 16.24 ± 14.47 % for T2. Using those pulses reduced the RMSE to 2.3 ± 8.4 % for T1 and 14.11 ± 13.46 % for T2. Nonetheless, even if the improvement is significant, RMSE are still too high for reliable quantification. CONCLUSION: In contrast to previous study using triggered MRF sequences using < 30° flip angles, large flip angle amplitudes led to better results for continuous cardiac MRF sequences. T2-preparation pulse can improve the accuracy of T2 estimation but lead to longer scan times.


Assuntos
Encéfalo , Imageamento por Ressonância Magnética , Humanos , Imageamento por Ressonância Magnética/métodos , Coração/diagnóstico por imagem , Espectroscopia de Ressonância Magnética , Imagens de Fantasmas , Processamento de Imagem Assistida por Computador/métodos
10.
Magn Reson Med ; 89(3): 1002-1015, 2023 03.
Artigo em Inglês | MEDLINE | ID: mdl-36336877

RESUMO

PURPOSE: Subject-tailored parallel transmission pulses for ultra-high fields body applications are typically calculated based on subject-specific B 1 + $$ {\mathrm{B}}_1^{+} $$ -maps of all transmit channels, which require lengthy adjustment times. This study investigates the feasibility of using deep learning to estimate complex, channel-wise, relative 2D B 1 + $$ {\mathrm{B}}_1^{+} $$ -maps from a single gradient echo localizer to overcome long calibration times. METHODS: 126 channel-wise, complex, relative 2D B 1 + $$ {\mathrm{B}}_1^{+} $$ -maps of the human heart from 44 subjects were acquired at 7T using a Cartesian, cardiac gradient-echo sequence obtained under breath-hold to create a library for network training and cross-validation. The deep learning predicted maps were qualitatively compared to the ground truth. Phase-only B 1 + $$ {\mathrm{B}}_1^{+} $$ -shimming was subsequently performed on the estimated B 1 + $$ {\mathrm{B}}_1^{+} $$ -maps for a region of interest covering the heart. The proposed network was applied at 7T to 3 unseen test subjects. RESULTS: The deep learning-based B 1 + $$ {\mathrm{B}}_1^{+} $$ -maps, derived in approximately 0.2 seconds, match the ground truth for the magnitude and phase. The static, phase-only pulse design performs best when maximizing the mean transmission efficiency. In-vivo application of the proposed network to unseen subjects demonstrates the feasibility of this approach: the network yields predicted B 1 + $$ {\mathrm{B}}_1^{+} $$ -maps comparable to the acquired ground truth and anatomical scans reflect the resulting B 1 + $$ {\mathrm{B}}_1^{+} $$ -pattern using the deep learning-based maps. CONCLUSION: The feasibility of estimating 2D relative B 1 + $$ {\mathrm{B}}_1^{+} $$ -maps from initial localizer scans of the human heart at 7T using deep learning is successfully demonstrated. Because the technique requires only sub-seconds to derive channel-wise B 1 + $$ {\mathrm{B}}_1^{+} $$ -maps, it offers high potential for advancing clinical body imaging at ultra-high fields.


Assuntos
Aprendizado Profundo , Imageamento por Ressonância Magnética , Humanos , Imageamento por Ressonância Magnética/métodos , Interpretação de Imagem Assistida por Computador/métodos , Coração/diagnóstico por imagem , Calibragem
11.
Magn Reson Med ; 88(5): 1978-1993, 2022 11.
Artigo em Inglês | MEDLINE | ID: mdl-35906900

RESUMO

PURPOSE: To simultaneously acquire spectroscopic signals from two MRS voxels using a multi-banded 2 spin-echo, full-intensity acquired localized (2SPECIAL) sequence, and to decompose the signal to their respective regions by a novel voxel-GRAPPA (vGRAPPA) decomposition approach for in vivo brain applications at 7 T. METHODS: A wideband, uniform rate, smooth truncation (WURST) multi-banded pulse was incorporated into SPECIAL to implement 2SPECIAL for simultaneous multi-voxel spectroscopy (sMVS). To decompose the acquired data, the voxel-GRAPPA decomposition algorithm is introduced, and its performance is compared to the SENSE-based decomposition. Furthermore, the limitations of two-voxel excitation concerning the multi-banded adiabatic inversion pulse, as well as of the combined B0 shim and B1 + adjustments, are evaluated. RESULTS: It was successfully shown that the 2SPECIAL sequence enables sMVS without a significant loss in SNR while reducing the total scan time by 21.6% compared to two consecutive acquisitions. The proposed voxel-GRAPPA algorithm properly reassigns the signal components to their respective origin region and shows no significant differences to the well-established SENSE-based algorithm in terms of leakage (both <10%) or Cramér-Rao lower bounds (CRLB) for in vivo applications, while not requiring the acquisition of additional sensitivity maps and thus decreasing motion sensitivity. CONCLUSION: The use of 2SPECIAL in combination with the novel voxel-GRAPPA decomposition technique allows a substantial reduction of measurement time compared to the consecutive acquisition of two single voxels without a significant decrease in spectral quality or metabolite quantification accuracy and thus provides a new option for multiple-voxel applications.


Assuntos
Algoritmos , Encéfalo , Encéfalo/diagnóstico por imagem , Encéfalo/metabolismo , Movimento (Física)
12.
Magn Reson Med ; 88(6): 2645-2661, 2022 12.
Artigo em Inglês | MEDLINE | ID: mdl-35906923

RESUMO

PURPOSE: To present electromagnetic simulation setups for detailed analyses of respiration's impact on B 1 + $$ {B}_1^{+} $$ and E-fields, local specific absorption rate (SAR) and associated safety-limits for 7T cardiac imaging. METHODS: Finite-difference time-domain electromagnetic field simulations were performed at five respiratory states using a breathing body model and a 16-element 7T body transceiver RF-coil array. B 1 + $$ {B}_1^{+} $$ and SAR are analyzed for fixed and moving coil configurations. SAR variations are investigated using phase/amplitude shimming considering (i) a local SAR-controlled mode (here SAR calculations consider RF amplitudes and phases) and (ii) a channel-wise power-controlled mode (SAR boundary calculation is independent of the channels' phases, only dependent on the channels' maximum amplitude). RESULTS: Respiration-induced variations of both B 1 + $$ {B}_1^{+} $$ amplitude and phase are observed. The flip angle homogeneity depends on the respiratory state used for B 1 + $$ {B}_1^{+} $$ shimming; best results were achieved for shimming on inhale and exhale simultaneously ( | Δ C V | < 35 % $$ \mid \Delta CV\mid <35\% $$ ). The results reflect that respiration impacts position and amplitude of the local SAR maximum. With the local-SAR-control mode, a safety factor of up to 1.4 is needed to accommodate for respiratory variations while the power control mode appears respiration-robust when the coil moves with respiration (SAR peak decrease: 9% exhale→inhale). Instead, a spatially fixed coil setup yields higher SAR variations with respiration. CONCLUSION: Respiratory motion does not only affect the B 1 + $$ {B}_1^{+} $$ distribution and hence the image contrast, but also location and magnitude of the peak spatial SAR. Therefore, respiration effects may need to be included in safety analyses of RF coils applied to the human thorax.


Assuntos
Campos Eletromagnéticos , Imageamento por Ressonância Magnética , Simulação por Computador , Humanos , Imageamento por Ressonância Magnética/métodos , Imagens de Fantasmas , Ondas de Rádio
13.
Magn Reson Med ; 87(6): 2862-2871, 2022 06.
Artigo em Inglês | MEDLINE | ID: mdl-35142400

RESUMO

PURPOSE: Human heart imaging at ultra-high fields is highly challenging because of respiratory motion-induced artefacts and spatially heterogeneous B1+ profiles. This work demonstrates that respiration resolved 3D B1+ -maps can be used with a dedicated tailored and universal parallel transmission (pTx) pulse design to compensate respiration related B1+ changes in subjects performing shallow and deep breathing (SB/DB). METHODS: Three-dimensional (3D) B1+ -maps of the thorax were acquired in 31 subjects under SB and in 15 subjects under SB and DB. Different universal and tailored non-selective pTx pulses were designed from non-respiration resolved (NRR) and respiration resolved (RR) reconstructions of the SB/DB B1+ -maps. The performance of all pulses was tested with RR-SB/DB B1+ -maps. Respiration-robust tailored and universal pulses were applied in vivo in 5 subjects at 7T in 3D gradient-echo free-breathing scans. RESULTS: All optimized pTx pulses performed well for SB. For DB, however, only the universal and the tailored respiration-robust pulses achieved homogeneous flip angles (FAs) in all subjects and across all respiration states, whereas the tailored respiration-specific pulses resulted in a higher FA variation. The respiration-robust universal pulse resulted in an average coefficient of variation in the FA maps of 12.6% compared to 8.2% achieved by tailored respiration-robust pulses. In vivo measurements at 7T demonstrate the benefits of using respiration-robust pulses for DB. CONCLUSION: Universal and tailored respiration-robust pTx pulses based on RR B1+ -maps are highly preferred to achieve 3D heart FA homogenization at 7T when subjects perform DB, whereas universal and tailored pulses based on NRR B1+ -maps are sufficient when subjects perform SB.


Assuntos
Encéfalo , Imageamento por Ressonância Magnética , Algoritmos , Artefatos , Técnicas de Imagem Cardíaca , Humanos , Imageamento por Ressonância Magnética/métodos , Respiração
14.
Magn Reson Med ; 87(6): 2621-2636, 2022 06.
Artigo em Inglês | MEDLINE | ID: mdl-35092090

RESUMO

PURPOSE: Respiratory motion-compensated (MC) 3D cardiac fat-water imaging at 7T. METHODS: Free-breathing bipolar 3D triple-echo gradient-recalled-echo (GRE) data with radial phase-encoding (RPE) trajectory were acquired in 11 healthy volunteers (7M\4F, 21-35 years, mean: 30 years) with a wide range of body mass index (BMI; 19.9-34.0 kg/m2 ) and volunteer tailored B1+ shimming. The bipolar-corrected triple-echo GRE-RPE data were binned into different respiratory phases (self-navigation) and were used for the estimation of non-rigid motion vector fields (MF) and respiratory resolved (RR) maps of the main magnetic field deviations (ΔB0 ). RR ΔB0 maps and MC ΔB0 maps were compared to a reference respiratory phase to assess respiration-induced changes. Subsequently, cardiac binned fat-water images were obtained using a model-based, respiratory motion-corrected image reconstruction. RESULTS: The 3D cardiac fat-water imaging at 7T was successfully demonstrated. Local respiration-induced frequency shifts in MC ΔB0 maps are small compared to the chemical shifts used in the multi-peak model. Compared to the reference exhale ΔB0 map these changes are in the order of 10 Hz on average. Cardiac binned MC fat-water reconstruction reduced respiration induced blurring in the fat-water images, and flow artifacts are reduced in the end-diastolic fat-water separated images. CONCLUSION: This work demonstrates the feasibility of 3D fat-water imaging at UHF for the entire human heart despite spatial and temporal B1+ and B0 variations, as well as respiratory and cardiac motion.


Assuntos
Imageamento por Ressonância Magnética , Água , Artefatos , Humanos , Imageamento Tridimensional , Movimento (Física) , Respiração
15.
J Magn Reson Imaging ; 55(1): 211-222, 2022 01.
Artigo em Inglês | MEDLINE | ID: mdl-34173297

RESUMO

BACKGROUND: Implementation of four-dimensional flow magnetic resonance (4D Flow MR) in clinical routine requires awareness of confounders. PURPOSE: To investigate inter-vendor comparability of 4D Flow MR derived aortic hemodynamic parameters, assess scan-rescan repeatability, and intra- and interobserver reproducibility. STUDY TYPE: Prospective multicenter study. POPULATION: Fifteen healthy volunteers (age 24.5 ± 5.3 years, 8 females). FIELD STRENGTH/SEQUENCE: 3 T, vendor-provided and clinically used 4D Flow MR sequences of each site. ASSESSMENT: Forward flow volume, peak velocity, average, and maximum wall shear stress (WSS) were assessed via nine planes (P1-P9) throughout the thoracic aorta by a single observer (AD, 2 years of experience). Inter-vendor comparability as well as scan-rescan, intra- and interobserver reproducibility were examined. STATISTICAL TESTS: Equivalence was tested setting the 95% confidence interval of intraobserver and scan-rescan difference as the limit of clinical acceptable disagreement. Intraclass correlation coefficient (ICC) and Bland-Altman plots were used for scan-rescan reproducibility and intra- and interobserver agreement. A P-value <0.05 was considered statistically significant. ICCs ≥ 0.75 indicated strong correlation (>0.9: excellent, 0.75-0.9: good). RESULTS: Ten volunteers finished the complete study successfully. 4D flow derived hemodynamic parameters between scanners of three different vendors are not equivalent exceeding the equivalence range. P3-P9 differed significantly between all three scanners for forward flow (59.1 ± 13.1 mL vs. 68.1 ± 12.0 mL vs. 55.4 ± 13.1 mL), maximum WSS (1842.0 ± 190.5 mPa vs. 1969.5 ± 398.7 mPa vs. 1500.6 ± 247.2 mPa), average WSS (1400.0 ± 149.3 mPa vs. 1322.6 ± 211.8 mPa vs. 1142.0 ± 198.5 mPa), and peak velocity between scanners I vs. III (114.7 ± 12.6 cm/s vs. 101.3 ± 15.6 cm/s). Overall, the plane location at the sinotubular junction (P1) presented most inter-vendor stability (forward: 78.5 ± 15.1 mL vs. 80.3 ± 15.4 mL vs. 79.5 ± 19.9 mL [P = 0.368]; peak: 126.4 ± 16.7 cm/s vs. 119.7 ± 13.6 cm/s vs. 111.2 ± 22.6 cm/s [P = 0.097]). Scan-rescan reproducibility and intra- and interobserver variability were good to excellent (ICC ≥ 0.8) with best agreement for forward flow (ICC ≥ 0.98). DATA CONCLUSION: The clinical protocol used at three different sites led to differences in hemodynamic parameters assessed by 4D flow. LEVEL OF EVIDENCE: 2 TECHNICAL EFFICACY STAGE: 2.


Assuntos
Hemodinâmica , Imageamento por Ressonância Magnética , Adulto , Humanos , Espectroscopia de Ressonância Magnética , Estudos Prospectivos , Reprodutibilidade dos Testes , Adulto Jovem
16.
Magn Reson Med ; 87(1): 70-84, 2022 01.
Artigo em Inglês | MEDLINE | ID: mdl-34399002

RESUMO

PURPOSE: MRI at ultra-high fields in the human body is highly challenging and requires lengthy calibration times to compensate for spatially heterogeneous B1+ profiles. This study investigates the feasibility of using pre-computed universal pulses for calibration-free homogeneous 3D flip angle distribution in the human heart at 7T. METHODS: Twenty-two channel-wise 3D B1+ data sets were acquired under free-breathing in 19 subjects to generate a library for an offline universal pulse (UP) design (group 1: 12 males [M] and 7 females [F], 21-66 years, 19.8-28.3 kg/m2 ). Three of these subjects (2M/1F, 21-33 years, 20.8-23.6 kg/m2 ) were re-scanned on different days. A 4kT-points UP optimized for the 22 channel-wise 3D B1+ data sets in group 1 (UP22-4kT) is proposed and applied at 7T in 9 new and unseen subjects (group 2: 4M/5F, 25-56 years, 19.5-35.3 kg/m2 ). Multiple tailored and universal static and dynamic parallel-transmit (pTx) pulses were designed and evaluated for different permutations of the B1+ data sets in group 1 and 2. RESULTS: The proposed UP22-4kT provides low B1+ variation in all subjects, seen and unseen, without severe signal drops. Experimental data at 7T acquired with UP22-4kT shows comparable image quality as data acquired with tailored-4kT pulses and demonstrates successful calibration-free pTx of the human heart. CONCLUSION: UP22-4kT allows for calibration-free homogeneous flip angle distributions across the human heart at 7T. Large inter-subject variations because of sex, age, and body mass index are well tolerated. The proposed universal pulse removes the need for lengthy (10-15 min) calibration scans and therefore has the potential to bring body imaging at 7T closer to the clinical application.


Assuntos
Algoritmos , Imageamento por Ressonância Magnética , Encéfalo , Calibragem , Feminino , Coração/diagnóstico por imagem , Humanos , Masculino , Respiração
17.
Magn Reson Med ; 87(1): 207-219, 2022 01.
Artigo em Inglês | MEDLINE | ID: mdl-34411335

RESUMO

PURPOSE: To improve pseudo-continuous arterial spin labeling (pcASL) at 7T by exploiting a hybrid homogeneity- and efficiency-optimized B1+ -shim with adapted gradient strength as well as background suppression. METHODS: The following three experiments were performed at 7T, each employing five volunteers: (1) A hybrid (ie, homogeneity-efficiency optimized) B1+ -shim was introduced and evaluated for variable-rate selective excitation pcASL labeling. Therefore, B1+ -maps in the V3 segment and time-of-flight images were acquired to identify the feeding arteries. For validation, a gradient-echo sequence was applied in circular polarized (CP) mode and with the hybrid B1+ -shim. Additionally, the gray matter (temporal) signal-to-noise ratio (tSNR) in pcASL perfusion images was evaluated. (2) Bloch simulations for the pcASL labeling were conducted and validated experimentally, with a focus on the slice-selective gradients. (3) Background suppression was added to the B1+ -shimmed, gradient-adapted 7T sequence and this was then compared to a matched sequence at 3T. RESULTS: The B1+ -shim improved the signal within the labeling plane (23.6%) and the SNR/tSNR increased (+11%/+11%) compared to its value in CP mode; however, the increase was not significant. In accordance with the simulations, the adapted gradients increased the tSNR (35%) and SNR (45%) significantly. Background suppression further improved the perfusion images at 7T, and this protocol performed as well as a resolution-matched protocol at 3T. CONCLUSION: The combination of the proposed hybrid B1+ -phase-shim with the adapted slice-selective gradients and background suppression shows great potential for improved pcASL labeling under suboptimal B1+ conditions at 7T.


Assuntos
Artérias , Encéfalo , Encéfalo/diagnóstico por imagem , Circulação Cerebrovascular , Substância Cinzenta/diagnóstico por imagem , Razão Sinal-Ruído , Marcadores de Spin
18.
Magn Reson Med ; 87(3): 1119-1135, 2022 03.
Artigo em Inglês | MEDLINE | ID: mdl-34783376

RESUMO

PURPOSE: To introduce a study design and statistical analysis framework to assess the repeatability, reproducibility, and minimal detectable changes (MDCs) of metabolite concentrations determined by in vivo MRS. METHODS: An unbalanced nested study design was chosen to acquire in vivo MRS data within different repeatability and reproducibility scenarios. A spin-echo, full-intensity acquired localized (SPECIAL) sequence was employed at 7 T utlizing three different inversion pulses: a hyperbolic secant (HS), a gradient offset independent adiabaticity (GOIA), and a wideband, uniform rate, smooth truncation (WURST) pulse. Metabolite concentrations, Cramér-Rao lower bounds (CRLBs) and coefficients of variation (CVs) were calculated. Both Bland-Altman analysis and a restricted maximum-likelihood estimation (REML) analysis were performed to estimate the different variance contributions of the repeatability and reproducibility of the measured concentration. A Bland-Altmann analysis of the spectral shape was performed to assess the variance of the spectral shape, independent of quantification model influences. RESULTS: For the used setup, minimal detectable changes of brain metabolite concentrations were found to be between 0.40 µmol/g and 2.23 µmol/g. CRLBs account for only 16 % to 74 % of the total variance of the metabolite concentrations. The application of gradient-modulated inversion pulses in SPECIAL led to slightly improved repeatability, but overall reproducibility appeared to be limited by differences in positioning, calibration, and other day-to-day variations throughout different sessions. CONCLUSION: A framework is introduced to estimate the precision of metabolite concentrations obtained by MRS in vivo, and the minimal detectable changes for 13 metabolite concentrations measured at 7 T using SPECIAL are obtained.


Assuntos
Encéfalo , Encéfalo/diagnóstico por imagem , Humanos , Espectroscopia de Ressonância Magnética , Reprodutibilidade dos Testes
19.
Magn Reson Med ; 86(6): 3201-3210, 2021 12.
Artigo em Inglês | MEDLINE | ID: mdl-34313340

RESUMO

PURPOSE: To develop a phase-contrast (PC) -based method for direct and unbiased quantification of the acceleration vector field by synchronization of the spatial and acceleration encoding time points. The proposed method explicitly aims at in-vitro applications, requiring high measurement accuracy, as well as the validation of clinically relevant acceleration-encoded sequences. METHODS: A velocity-encoded sequence with synchronized encoding (SYNC SPI) was modified to allow direct acceleration mapping by replacing the bipolar encoding gradients with tripolar gradient waveforms. The proposed method was validated in two in-vitro flow cases: a rotation and a stenosis phantom. The thereby obtained velocity and acceleration vector fields were quantitatively compared to those acquired with conventional PC methods, as well as to theoretical data. RESULTS: The rotation phantom study revealed a systematic bias of the conventional PC acceleration mapping method that resulted in an average pixel-wise relative angle between the measured and theoretical vector field of (7.8 ± 3.2)°, which was reduced to (-0.4 ± 2.7)° for the proposed SYNC SPI method. Furthermore, flow features in the stenosis phantom were displaced by up to 10 mm in the conventional PC data compared with the acceleration-encoded SYNC SPI data. CONCLUSIONS: This work successfully demonstrates a highly accurate method for direct acceleration mapping. It thus complements the existing velocity-encoded SYNC SPI method to enable the direct and unbiased quantification of both the velocity and acceleration vector field for in vitro studies. Hence, this method can be used for the validation of conventional acceleration-encoded PC methods applicable in-vivo.


Assuntos
Aceleração , Imageamento por Ressonância Magnética , Velocidade do Fluxo Sanguíneo , Microscopia de Contraste de Fase , Imagens de Fantasmas
20.
Magn Reson Med ; 86(5): 2412-2425, 2021 11.
Artigo em Inglês | MEDLINE | ID: mdl-34061397

RESUMO

PURPOSE: To develop a framework for 3D sodium (23 Na) MR fingerprinting (MRF), based on irreducible spherical tensor operators with tailored flip angle (FA) pattern and time-efficient data acquisition for simultaneous quantification of T1 , T2l∗ , T2s∗ , and T2∗ in addition to ΔB0 . METHODS: 23 Na-MRF was implemented in a 3D sequence and irreducible spherical tensor operators were exploited in the simulations. Furthermore, the Cramér Rao lower bound was used to optimize the flip angle pattern. A combination of single and double echo readouts was implemented to increase the readout efficiency. A study was conducted to compare results in a multicompartment phantom acquired with MRF and reference methods. Finally, the relaxation times in the human brain were measured in four healthy volunteers. RESULTS: Phantom experiments revealed a mean difference of 1.0% between relaxation times acquired with MRF and results determined with the reference methods. Simultaneous quantification of the longitudinal and transverse relaxation times in the human brain was possible within 32 min using 3D 23 Na-MRF with a nominal resolution of (5 mm)3 . In vivo measurements in four volunteers yielded average relaxation times of: T1,brain = (35.0 ± 3.2) ms, T2l,brain∗ = (29.3 ± 3.8) ms and T2s,brain∗ = (5.5 ± 1.3) ms in brain tissue, whereas T1,CSF = (61.9 ± 2.8) ms and T2,CSF∗ = (46.3 ± 4.5) ms was found in cerebrospinal fluid. CONCLUSION: The feasibility of in vivo 3D relaxometric sodium mapping within roughly ½ h is demonstrated using MRF in the human brain, moving sodium relaxometric mapping toward clinically relevant measurement times.


Assuntos
Imageamento por Ressonância Magnética , Sódio , Encéfalo/diagnóstico por imagem , Humanos , Processamento de Imagem Assistida por Computador , Espectroscopia de Ressonância Magnética , Imagens de Fantasmas