Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros











Base de dados
Intervalo de ano de publicação
1.
Sci Rep ; 14(1): 3560, 2024 02 12.
Artigo em Inglês | MEDLINE | ID: mdl-38347154

RESUMO

We sought to compare the retinal thickness measurements collected using different optical coherence tomography (OCT) devices. This prospective study included 21 healthy cases, and the retinal thickness was measured using the PLEX Elite (Carl Zeiss Meditec, Dublin, California, USA), DRI OCT-1 Atlantis (Topcon Corp, Tokyo, Japan), Cirrus 5000 HD-OCT (Carl Zeiss Meditec), and Spectralis OCT (Heidelberg Engineering, Heidelberg, Germany), respectively. The mean central retinal thickness (CRT) and mean retinal thickness of the Early Treatment of Diabetic Retinopathy Study (ETDRS) area were compared. The CRT varied significantly among the different OCT devices (P < 0.001). Post-hoc analysis revealed that the CRT measured using PLEX Elite (278.95 ± 20.04 µm) and Spectralis (271.86 ± 17.92 µm) were similar, and both were greater than the CRT measurements of DRI OCT-1 (239.57 ± 21.06 µm) and Cirrus (256.76 ± 17.82 µm). Additionally, the mean retinal thickness in each ETDRS area showed significant differences among the four devices (all P < 0.001). The mean retinal thickness measured varied according to the device used, and this needs to be considered when comparing retinal thickness measurements taken with different devices.


Assuntos
Retinopatia Diabética , Tomografia de Coerência Óptica , Humanos , Estudos Prospectivos , Reprodutibilidade dos Testes , Retina/diagnóstico por imagem , Retinopatia Diabética/diagnóstico por imagem
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA