Artigo em Inglês | WPRIM (Pacífico Ocidental) | ID: wpr-888486

RESUMO

OBJECTIVES@#To evaluate the accuracy and safety of measurements of transcutaneous carbon dioxide partial pressure (TcPCO@*METHODS@#A total of 45 very low birth weight infants were enrolled. TcPCO@*RESULTS@#There was no significant difference in TcPCO@*CONCLUSIONS@#Lower electrode temperatures (38-41℃) can accurately measure blood carbon dioxide partial pressure in very low birth weight infants, and thus can be used to replace the electrode temperature of 42°C. Transcutaneous measurements at the lower electrode temperatures may be helpful for understanding the changing trend of blood oxygen partial pressure.


Assuntos
Humanos , Lactente , Recém-Nascido , Monitorização Transcutânea dos Gases Sanguíneos , Dióxido de Carbono , Eletrodos , Recém-Nascido de muito Baixo Peso , Oxigênio , Pressão Parcial , Temperatura