Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
J Am Soc Mass Spectrom ; 27(11): 1789-1795, 2016 11.
Artigo em Inglês | MEDLINE | ID: mdl-27527096

RESUMO

As a reagent gas for positive- and negative-mode chemical ionization mass spectrometry (CI-MS), isobutane (i-C4H10) produces superior analyte signal abundance to methane. Isobutane has never been widely adopted for CI-MS because it fouls the ion source more rapidly and produces positive CI spectra that are more strongly dependent on reagent gas pressure compared with methane. Isobutane was diluted to various concentrations in argon for use as a reagent gas with an unmodified commercial gas chromatograph-mass spectrometer. Analyte spectra were directly compared using methane, isobutane, and isobutane/argon mixtures. A mixture of 10% i-C4H10 in argon produced twice the positive-mode analyte signal of methane, equal to pure isobutane, and reduced spectral dependence on reagent gas pressure. Electron capture negative chemical ionization using 1% i-C4H10 in argon tripled analyte signal compared with methane and was reproducible, unlike pure isobutane. The operative lifetime of the ion source using isobutane/argon mixtures was extended exponentially compared with pure isobutane, producing stable and reproducible CI signal throughout. By diluting the reagent gas in an inert buffer gas, isobutane CI-MS experiments were made as practical to use as methane CI-MS experiments but with superior analytical performance. Graphical Abstract ᅟ.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...