1.
Clin Biochem ; 44(13): 1128-1136, 2011 Sep.
Artigo em Inglês | MEDLINE | ID: mdl-21704611

RESUMO

OBJECTIVES: We examined analytical characteristics of new CA 15-3, CA 19-9, CA 125 II, Carcinoembryonic Antigen (CEA), and Alpha-Fetoprotein (AFP) assays on the Dimension Vista® System. DESIGN AND METHODS: Imprecision studies used CLSI-EP5-A2, Limit of Blank and Limit of Detection used CLSI-EP17 and measurement ranges were determined. Method comparisons were evaluated with Passing-Bablok, least-squares regression and residual plots. Reference intervals were determined and valid specimen types, lot-to-lot variability and sample storage stability were defined. Clinical monitoring patterns for each tumor marker in patients were examined. RESULTS: Reproducibility for each method was <6.5%. Limits of Blank and Detection were low. Comparisons between methods showed slopes ranging from 0.89 to 1.32 with low y-intercepts and scatter. Minimal lot-to-lot variability was documented; serum/plasma specimens provide valid results; sample stability at -70°C was >9months. Clinical monitoring patterns correlated with established methods in >89% of cases. CONCLUSIONS: Measurement of CA 15-3, CA 19-9, CA 125 II, CEA and AFP on the Dimension Vista® System is an attractive alternative.


Assuntos
Antígenos Glicosídicos Associados a Tumores/análise , Biomarcadores Tumorais/análise , Técnicas e Procedimentos Diagnósticos/instrumentação , Neoplasias/diagnóstico , alfa-Fetoproteínas/análise , Antígeno Ca-125/análise , Antígeno CA-19-9/análise , Antígeno Carcinoembrionário/análise , Técnicas e Procedimentos Diagnósticos/normas , Humanos , Limite de Detecção , Mucina-1/análise , Reprodutibilidade dos Testes , Sensibilidade e Especificidade
3.
Clin Biochem ; 42(13-14): 1444-51, 2009 Sep.
Artigo em Inglês | MEDLINE | ID: mdl-19523464

RESUMO

OBJECTIVES: Performance characteristics of the LOCI cTnI, CK-MB, MYO, NTproBNP and hsCRP methods on the Dimension Vista System were evaluated. DESIGN AND METHODS: Imprecision (following CLSI EP05-A2 guidelines), limit of quantitation (cTnI), limit of blank, linearity on dilution, serum versus plasma matrix studies (cTnI), and method comparison studies were conducted. RESULTS: Method imprecision of 1.8 to 9.7% (cTnI), 1.8 to 5.7% (CK-MB), 2.1 to 2.2% (MYO), 1.6 to 3.3% (NTproBNP), and 3.5 to 4.2% (hsCRP) were demonstrated. The manufacturer's claimed imprecision, detection limits and upper measurement limits were met. Limit of Quantitation was 0.040 ng/mL for the cTnI assay. Agreement of serum and plasma values for cTnI (r=0.99) was shown. Method comparison study results were acceptable. CONCLUSIONS: The Dimension Vista cTnI, CK-MB, MYO, NTproBNP, and hsCRP methods demonstrate acceptable performance characteristics for use as an aid in the diagnosis and risk assessment of patients presenting with suspected acute coronary syndromes.


Assuntos
Síndrome Coronariana Aguda/sangue , Biomarcadores/sangue , Doenças Cardiovasculares/sangue , Imunoensaio/métodos , Síndrome Coronariana Aguda/diagnóstico , Proteína C-Reativa/análise , Doenças Cardiovasculares/diagnóstico , Creatina Quinase Forma MB/sangue , Humanos , Imunoensaio/instrumentação , Mioglobina/sangue , Peptídeo Natriurético Encefálico/sangue , Fragmentos de Peptídeos/sangue , Reprodutibilidade dos Testes , Medição de Risco , Sensibilidade e Especificidade , Troponina I/sangue