RESUMO
BACKGROUND: Self-monitoring of blood glucose is a cornerstone of diabetes management. The aim of this study was to evaluate the analytical quality and the ease of use of the Accu-Chek Mobile, a new glucose monitoring system designed for capillary blood testing by diabetic patients. MATERIALS AND METHODS: The performance of the Accu-Chek Mobile was evaluated both in the hands of a scientist and of diabetes patients. The designated comparative method was a hexokinase-based laboratory method (Architect ci8200). Diabetics (N = 88) with previous experience of self-testing were recruited for the study. Patient samples, containing glucose in concentrations mainly between Ë4 and Ë20 mmol/L, were analyzed in duplicates both on the Accu-Chek Mobile and with the comparative method. The patients answered a questionnaire about the ease of use of the meter. RESULTS: The meter yields reproducible readings, with an imprecision CV <5% as required by the American Diabetes Association (ADA). Of the glucose concentrations obtained by both the scientist and the patients, more than 95% of the individual results were within ± 20% of the comparative method, meeting the ISO 15197 accuracy goal, but not the stricter ± 10% ADA goal. CONCLUSION: Accu-Chek Mobile is a user-friendly glucometer that in a normo- and hyperglycemic range fulfils the ISO 15197 accuracy requirement, also in the hands of diabetes patients.