1.
Ann Biol Clin (Paris)
; 62(6): 691-4, 2004.
Artigo
em Francês
| MEDLINE
| ID: mdl-15563428
RESUMO
We evaluated an automated method for measuring low-density lipoprotein cholesterol (LDL-cholesterol) with Integra 800 (Roche Diagnostics). This method uses detergents which show selectivity for the different lipoproteins. Our evaluation has shown a within run imprecision of 2% and a between-assay imprecision of 1.2%. This method was not perturbed by haemoglobin, bilirubin or lipids. However, the reagents remain expensive: we will use it when calculation or interpretation of lipid results are not possible.