RESUMO
BACKGROUND: The aim was to investigate to what extent the quantification of myocardial blood flow (MBF) from dynamic 13 N-NH3 positron emission tomography (PET) images is affected by time frame schemes, time-of-flight (ToF), reconstruction algorithms, blood pool volume of interest (VOI) locations and compartment models in patients with suspected chronic coronary syndrome. METHODS: A standard MBF value was determined from 25 patients' rest/stress 13 N-NH3 PET/CT images reconstructed with ordered subset expectation maximization (OSEM), 5 s time frame for the first frames without ToF, subsequently analyzed using a basal VOI and the deGrado compartment model. MBFs calculated using 2 or 10 s for the first frames, ToF, block-sequential regularized expectation maximization (BSREM), apical or large VOI, Hutchins or Krivokapich compartment models were compared to MBFstandard in Bland-Altman plots (bias ± SD). RESULTS: Good agreement in global rest/stress MBF (mL/min/g) was found when changing the time frame scheme or reconstruction algorithm (MBFstandard vs. MBF2s : -0.02 ± 0.06; MBF10s : 0.01 ± 0.07; MBFBSREM : 0.01 ± 0.07), while a lower level of agreement was found when altering the other factors (MBFstandard vs. MBFToF : -0.07 ± 0.10; MBFapical VOI : -0.27 ± 0.25; MBFlarge VOI : -0.11 ± 0.10; MBFHutchins : -0.08 ± 0.10; MBFKrivokapich : -0.47 ± 0.50). CONCLUSIONS: Quantification of MBF from 13 N-NH3 PET images is more affected by choice of compartment models, ToF and blood pool VOIs than by different time frame schemes and reconstruction algorithms.