J Clin Monit Comput ; 32(6): 1155-1156, 2018 12.
Artigo em Inglês | MEDLINE | ID: mdl-29488044

RESUMO

Distinguishing a monitoring artifact requires expertise and adeptness. This can be practically challenging during the course of an anesthetic. We report a case, wherein we experienced episodes of aberrant pulse-oximeter values suggestive of desaturation with normal waveforms, occurring during a particular sequence of magnetic resonance imaging (MRI) performed under general anesthesia, which in fact was an artifact induced by the 3 T MRI during the diffusion tensor imaging sequence.


Assuntos
Hipóxia/diagnóstico , Imageamento por Ressonância Magnética , Monitorização Intraoperatória , Oximetria , Anestesia Geral , Artefatos , Encéfalo/diagnóstico por imagem , Criança , Reações Falso-Positivas , Humanos , Imageamento por Ressonância Magnética/efeitos adversos , Imageamento por Ressonância Magnética/estatística & dados numéricos , Masculino , Monitorização Intraoperatória/efeitos adversos , Monitorização Intraoperatória/estatística & dados numéricos , Oximetria/efeitos adversos , Oximetria/estatística & dados numéricos , Convulsões/diagnóstico por imagem