1.
Eye (Lond) ; 23(7): 1582-8, 2009 Jul.
Artigo em Inglês | MEDLINE | ID: mdl-19407845

RESUMO

AIMS: To compare intraocular pressure (IOP) readings between Tono-Pen tonometry and GAT, between noncontact tonometry (NCT) and GAT, and between dynamic contour tonometry (DCT) and Goldmann applanation tonometry (GAT). The correlation between IOP reading and possible confounder was identified. METHODS: This observational cross-sectional study enrolled sixty-two healthy subjects. All IOP and ocular pulse amplitude (OPA) measurements were taken by a single ophthalmologist; mean keratometric power (MK), central corneal thickness (CCT), and lens thickness (LT) were measured by a single experienced technician. RESULTS: Stepwise multiple regression analysis indicated that GAT (P=0.017) and DCT (P=0.002) readings correlated positively with MK; GAT, NCT, and Tono-Pen readings correlated positively with CCT (P<0.05); NCT (P=0.035), and DCT (P=0.016) readings correlated negatively with LT; GAT (P=0.006) and Tono-Pen (P=0.009) readings correlated positively with OPA. CONCLUSIONS: The K, CCT, LT, and OPA are confounders in tonometry readings.


Assuntos
Pressão Intraocular/fisiologia , Tonometria Ocular/métodos , Adulto , Idoso , Córnea/fisiologia , Estudos Transversais , Feminino , Humanos , Masculino , Pessoa de Meia-Idade , Análise de Regressão , Tonometria Ocular/instrumentação , Adulto Jovem