Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 6 de 6
Filtrar
Mais filtros











Base de dados
Intervalo de ano de publicação
1.
Life Sci Space Res (Amst) ; 7: 90-9, 2015 Nov.
Artigo em Inglês | MEDLINE | ID: mdl-26553642

RESUMO

In order to define the ranges of relevant neutron energies for the purposes of measurement and dosimetry in space, we have performed a series of Monte Carlo transport model calculations that predict the neutron field created by Galactic Cosmic Ray interactions inside a variety of simple shielding configurations. These predictions indicate that a significant fraction of the neutron fluence and neutron effective dose lies in the region above 20 MeV up to several hundred MeV. These results are consistent over thicknesses of shielding that range from very thin (2.7 g/cm(2)) to thick (54 g/cm(2)), and over both shielding materials considered (aluminum and water). In addition to these results, we have also investigated whether simplified Galactic Cosmic Ray source terms can yield predictions that are equivalent to simulations run with a full GCR source term. We found that a source using a GCR proton and helium spectrum together with a scaled oxygen spectrum yielded nearly identical results to a full GCR spectrum, and that the scaling factor used for the oxygen spectrum was independent of shielding material and thickness. Good results were also obtained using a GCR proton spectrum together with a scaled helium spectrum, with the helium scaling factor also independent of shielding material and thickness. Using a proton spectrum alone was unable to reproduce the full GCR results.


Assuntos
Nêutrons , Radiação Cósmica , Modelos Teóricos , Método de Monte Carlo , Prótons , Doses de Radiação , Proteção Radiológica , Radiometria , Voo Espacial
2.
Appl Radiat Isot ; 69(12): 1897-900, 2011 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-21605978

RESUMO

Various quality control (QC) and quality assurance (QA) procedures of the boron neutron capture therapy (BNCT) beam at the Tsing Hua Open-pool Reactor (THOR) are established to ensure beam availability and quality. The QC/QA methods mainly employ foil activation and paired ionization chambers, respectively, for beam intensity check and dose assessment. Beam intensity is monitored on-line by using three dead-time corrected fission chambers. In addition to the periodic QC/QA activities regarding beam quality and the monitoring system, the quick QC/QA performed in an all-in-one phantom will be executed less than 70 min before the clinical treatment to guarantee beam quality. The QC/QA procedures have been gradually established and the actual performance satisfied the preset criteria defined for the BNCT facility at THOR.


Assuntos
Terapia por Captura de Nêutron de Boro , Controle de Qualidade
3.
Appl Radiat Isot ; 69(12): 1915-7, 2011 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-21570854

RESUMO

The GAFCHROMIC(®) EBT2 dosimetry film has been studied as a rapid QC/QA tool for 2D dose profile mapping in the BNCT beam at THOR. The pixel values of the EBT2 film image were converted to the 2D dose profile using a dose calibration curve obtained by 6-MV X-ray. The reproducibility of the 2D dose profile measured using the EBT2 film in the PMMA phantom was preliminarily found to be acceptable with uncertainties within about ±2 to ±3.5%. It is found that the EBT2 measured dose profile consisted of both gamma-ray components and neutron contributions. Therefore, the dose profile measured using the EBT2 film is significantly different from the neutron flux profile measured using the indirect neutron radiography method. Further study of the influence of neutrons to the response of the EBT2 film is indispensible for the absolute dose profile determination in a BNCT beam.


Assuntos
Terapia por Captura de Nêutron de Boro , Controle de Qualidade , Calibragem , Reprodutibilidade dos Testes
4.
Appl Radiat Isot ; 69(12): 1892-6, 2011 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-21570855

RESUMO

This paper aims to evaluate the performance of the source description of the THOR BNCT beam via different measurement techniques in different phantoms. The measurement included (1) the absolute reaction rate measurement of a set of triple activation foils, (2) the neutron and gamma-ray dose rates measured using the paired ionization chamber method, and (3) the relative reaction rate distributions obtained using the indirect neutron radiography. Three source descriptions, THOR-Y09, surface source file RSSA, and THOR-50C, were tested. The comparison results concluded that THOR-Y09 is a well-tested source description not only for neutron components, but also for gamma-ray component.


Assuntos
Terapia por Captura de Nêutron de Boro , Imagens de Fantasmas
5.
Appl Radiat Isot ; 67(7-8 Suppl): S137-40, 2009 Jul.
Artigo em Inglês | MEDLINE | ID: mdl-19409798

RESUMO

This paper aims to measure the spectra of HB11 (high flux reactor, HFR) and the Tsing Hua open-pool reactor (THOR) boron neutron capture therapy (BNCT) beams by multiple activation foils. The self-shielding corrections were made with the aid of MCNP calculations. The initial spectra were adjusted by a sophisticated process named coarse-scaling adjustment using SAND-EX, which can adjust a given coarse-group spectrum into a fine-group structure, i.e. 640 groups, with excellent continuity. The epithermal neutron flux of the THOR beam is about three times of HB11. The thermal neutron flux, boron and gold reaction rates along the central axis of a PMMA phantom are calculated for both adjusted spectra for comparison.


Assuntos
Terapia por Captura de Nêutron de Boro/estatística & dados numéricos , Nêutrons Rápidos/uso terapêutico , Terapia por Captura de Nêutron de Boro/instrumentação , Humanos , Reatores Nucleares , Imagens de Fantasmas/estatística & dados numéricos , Polimetil Metacrilato , Planejamento da Radioterapia Assistida por Computador/estatística & dados numéricos , Taiwan
6.
Appl Radiat Isot ; 67(7-8 Suppl): S190-4, 2009 Jul.
Artigo em Inglês | MEDLINE | ID: mdl-19394242

RESUMO

This study aims to measure the two-dimensional (2D) neutron spatial distribution inside a cubic and a head PMMA phantom for the purpose of further comparison with the treatment planning. The measurements were made by using the indirect neutron radiography (INR), which utilized a thin copper foil and the imaging plate. The developed image provides satisfactory spatial resolution and very low statistical error (< 1%). As to the time cost, the whole procedure normally takes less than 3 h. The result shows that the indirect neutron radiography can be a quick and reliable method to provide a 2D neutron spatial distribution inside a phantom.


Assuntos
Terapia por Captura de Nêutron de Boro/estatística & dados numéricos , Nêutrons Rápidos/uso terapêutico , Imagens de Fantasmas/estatística & dados numéricos , Planejamento da Radioterapia Assistida por Computador/estatística & dados numéricos , Terapia por Captura de Nêutron de Boro/instrumentação , Cobre , Cabeça , Humanos , Transferência Linear de Energia , Neoplasias/radioterapia , Polimetil Metacrilato , Taiwan
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA