1.
Talanta ; 72(1): 269-76, 2007 Apr 15.
Artigo em Inglês | MEDLINE | ID: mdl-19071613

RESUMO

A liquid chromatography-tandem mass spectrometry method for the determination and confirmation of nine quinolones was optimised and validated according to Commission Decision 2002/657/EC. Analytes were extracted from veal muscle with water and extracts purified with 96-well plates Oasis HLB cartridges. Separation was carried out in a silica-based C(18) column (50mmx2.1mm) with mobile phases consisting of water/acetonitrile mixtures containing acetic acid. Linear calibration curves in the ranges 4-400 and 50-800ngg(-1), with correlation coefficients at least 0.995, were obtained for all the analytes. At concentration levels above 10ngg(-1), quantification errors were lower than 10% and repeatability and within-laboratory reproducibility standard deviations below 6% and 10%, respectively. Decision limits and detection capabilities are reported.