RESUMO
OBJECTIVE@#A3 intertrochanteric fracture is an extremely unstable fracture, which is often treated with intramedullary nail, but the implant failure is common due to the posterior medial fragment cannot be reconstructed. A new medial sustainable nail (MSN-Ⅱ) which can reconstruct the femoral medial support by sustainable screw was introduced in this study. The mechanical effect was verified by biomechanical experiment.@*METHODS@#The loss medial support model of intertrochanteric fracture (A3) was made by artificial Sawbones model, fixed with MSN-Ⅱ and PFNA-Ⅱ, underwent axial loading and axial failure tests. The axial stiffness, yield load, displacement of head-neck fragment and torsional angle of fracture site of these nails were recorded and compared for biomechanical differences. The effect of early reconstruction of medial support with MSN-Ⅱ was determined.@*RESULTS@#The axial stiffness, yield load, the displacement of head and neck fragment when the axial load was 1 800 N and torsional angle of the fracture site after the axial failure test of MSN-Ⅱ were (222.76 ±62.46) N /mm, (4 241.71 ±847.42) N, (11.51 ±0.62) mm, (1.71 ±0.10)° respectively, while the PFNA -Ⅱ was (184.58±40.59) N /mm, (3 058.76±379.63) N, (16.15±1.36) mm, (2.52±0.26)°respectively. The difference between the two groups was statistically significant.@*CONCLUSION@#The axial stiffness of MSN-Ⅱ is better than that of PFNA-Ⅱ. The MSN-Ⅱ can bear more loads when fixed A3.3 intertrochanteric fracture and has greater axial and rotational stability. It is an effective means to reconstruct the medial support of A3 intertrochanteric fracture.