Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Blood Purif ; 50(6): 959-967, 2021.
Artigo em Inglês | MEDLINE | ID: mdl-33789265

RESUMO

BACKGROUND: By inhibiting the adsorption of protein and platelets, surface-modifying macromolecules (SMMs) may improve the hemocompatibility of hemodialyzers. This trial aims to assess the performance and safety of a novel dialyzer with a fluorinated polyurethane SMM, Endexo™. METHODS: This prospective, sequential, multicenter, open-label study (NCT03536663) was designed to meet regulatory requirements for clinical testing of new hemodialyzers, including assessment of the in vivo ultrafiltration coefficient (Kuf). Adults prescribed thrice-weekly hemodialysis were eligible for enrollment. After completing 12 hemodialysis sessions with an Optiflux® F160NR dialyzer, patients received 38 sessions with the dialyzer with Endexo. Evaluated parameters included the in vivo Kuf of the dialyzer with Endexo extent of removal of urea, albumin, and ß2-microglobulin (ß2M), as well as complement activation. RESULTS: Twenty-three patients received 268 hemodialysis treatments during the Optiflux period, and 18 patients received 664 hemodialysis treatments during the Endexo period. Three serious adverse events were reported, and none of them were considered device related. No overt complement activation was observed with either dialyzer. Both dialyzers were associated with comparable mean increases in serum albumin levels from pre- to posthemodialysis (Optiflux: 7.9%; Endexo: 8.0%). These increases can be viewed in the context of a mean increase in hemoglobin of approximately 5% and a mean ultrafiltration volume removed of approximately 2.2 L. The corrected mean ß2M removal rate was 47% higher during the Endexo period (67.73%). Mean treatment times (208 vs. 205 min), blood flow rates (447.7 vs. 447.5 mL/min), dialysate flow rates (698.5 vs. 698.0 mL/min), urea reduction ratio (82 vs. 81%), and spKt/V (2.1 vs. 1.9) were comparable for the Endexo and Optiflux periods, respectively. The mean (SD) Kuf was 15.85 (10.33) mL/h/mm Hg during the first use of the dialyzer with Endexo (primary endpoint) and 16.36 (9.92) mL/h/mm Hg across the Endexo period. CONCLUSIONS: The safety of the novel dialyzer with Endexo was generally comparable to the Optiflux dialyzer, while exhibiting a higher ß2M removal rate.


Assuntos
Materiais Biocompatíveis/química , Falência Renal Crônica/terapia , Poliuretanos/química , Diálise Renal/instrumentação , Adulto , Idoso , Idoso de 80 Anos ou mais , Materiais Biocompatíveis/efeitos adversos , Feminino , Halogenação , Humanos , Falência Renal Crônica/sangue , Masculino , Membranas Artificiais , Pessoa de Meia-Idade , Poliuretanos/efeitos adversos , Estudos Prospectivos , Diálise Renal/efeitos adversos , Albumina Sérica/análise , Albumina Sérica/isolamento & purificação , Ureia/sangue , Ureia/isolamento & purificação , Microglobulina beta-2/sangue , Microglobulina beta-2/isolamento & purificação
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...