Blood Press Monit ; 2024 Jul 30.
Artigo em Inglês | MEDLINE | ID: mdl-39105269

RESUMO

The aim of this study was to evaluate the accuracy of the JAMR upper-arm blood pressure monitor B23 in the general population according to the AAMI/ESH/ISO Universal Standard (ISO 81060-2 : 2018/AMD 1 : 2020). The study recruited participants who met the criteria of the AAMI/ESH/ISO Universal Standard in terms of their number, sex, age, limb size, and blood pressure (BP) distribution. The study involved measuring BP, including both SBP and DBP, using both the test device and a standard mercury sphygmomanometer in sequential measurements. Of 90 participants, 85 qualified participants were analyzed. A total of 255 sets of comparison data (three sets for each subject) were obtained and analyzed. For the validation criterion 1, the mean ±â€…SD of the differences between the JAMR B23 and mercury sphygmomanometer BP readings was -0.24 ±â€…6.52/-2.67 ±â€…5.6 mmHg (SBP/DBP). For criterion 2, the SD of the averaged BP (SBP/DBP) differences between the JAMR B23 and reference BP (SBP/DBP) per participant was 5.61/5.13 mmHg (the requirement was ≤6.95/6.43 mmHg by calculation). The JAMR B23 passed all the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-2 : 2018/AMD 1 : 2020) and can be recommended for clinical and self/home use in the general population.

2.
Blood Press Monit ; 28(3): 163-166, 2023 Jun 01.
Artigo em Inglês | MEDLINE | ID: mdl-37016972

RESUMO

To validate the JAMR F1701T (arm type) blood pressure (BP) monitor according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018). A total of 90 subjects (male 60 and female 30) were recruited to fulfill the criteria of the AAMI/ESH/ISO Universal Standard (the number, gender, age, limber size, and BP distribution), and sequential measurements of BP, including both SBP and DBP were obtained using the test device and the standard mercury sphygmomanometer. A total of 270 sets of comparison data (three sets of each subject) were obtained and analyzed. According to the validation criterion 1 of ISO 81060-2:2018, the mean ± SD of the differences between the JAMR F1701T and mercury sphygmomanometer BP (systolic/diastolic) readings was 2.06 ± 6.83/-4.84 ± 5.23 mmHg. For criterion 2, the SD of the averaged BP (systolic/diastolic) differences between the JAMR F1701 and reference BP (systolic/diastolic) per participant was 5.62/4.39 mmHg (the requirement was ≤6.43/5.01 mmHg by calculation). The JAMR F1701T met all the requirements of the ISO 81060-2:2018, and can be recommended for clinical and self/home use.


Assuntos
Hipertensão , Mercúrio , Humanos , Masculino , Feminino , Determinação da Pressão Arterial , Pressão Sanguínea , Braço , Hipertensão/diagnóstico , Monitores de Pressão Arterial
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA