Your browser doesn't support javascript.
loading
Avaliação da resistência à torção de mini-implantes após uso clínico e processo de reciclagem / Assessment of mini-implants resistance to torsion after clinical use and recycling process
Dezordi, Rodnei Luis; Vieira, Bruno da Silva; Santos, Sandro Henrique Batista; Cançado, Rodrigo Hermont; Valarelli, Fabrício Pinelli; Freitas, Karina Maria Salvatore.
Afiliação
  • Dezordi, Rodnei Luis; UNINGÁ. BR
  • Vieira, Bruno da Silva; USP. FOB. BR
  • Santos, Sandro Henrique Batista; UNINGÁ. BR
  • Cançado, Rodrigo Hermont; UFMG. BR
  • Valarelli, Fabrício Pinelli; UNINGÁ. BR
  • Freitas, Karina Maria Salvatore; UNINGÁ. BR
Ortho Sci., Orthod. sci. pract ; 15(58): 68-72, 2022. tab, ilus
Artigo em Português | BBO - Odontologia | ID: biblio-1380317
Biblioteca responsável: BR243.1
RESUMO
Resumo

Objetivo:

O objetivo do presente estudo foi avaliar a influência do uso clínico e do processo de reciclagem nos valores de torção e na resistência mecânica à fratura dos mini-implantes. Material e

métodos:

Trinta e quatro mini-implantes autoperfurantes de titânio (Ti-6AI-4V) de uma mesma marca comercial com 1,5 mm de diâmetro, 8,0 mm de comprimento e perfil transmucoso de 2,0 mm, que foram selecionados e divididos em dois grupos Grupo 1 foi composto de 24 mini-implantes inseridos na maxila e mandíbula de 17 pacientes sob tratamento ortodôntico e, que após 4 meses foram removidos e submetidos aos processos de reciclagem, incluindo limpeza e esterilização. O grupo 2 (controle) foi composto por 10 mini-implantes, conforme comercializados pelo fabricante. Os valores de torque de inserção e o torque máximo no momento da fratura dos mini-implantes foram realizados na máquina de teste universal (EMIC). Para comparação dos dados entre os grupos, foi realizado o t independente (p>0,05).

Resultados:

Não houve diferença estatisticamente significativa no teste de ensaio de resistência à torção entre os grupos analisados.

Conclusão:

Pode-se concluir que o processo de reciclagem dos mini-implantes não influenciou nos valores de resistência à torção, mantendo-se dentro do limite recomendado pelo fabricante para seu uso clínico.(AU)
ABSTRACT
Abstract

Objective:

The objective of the present study was to evaluate the influence of clinical use and the recycling process on torsion values and on the mechanical resistance to fracture of mini-implants. Material and

methods:

Thirty-four self-drilling titanium (Ti-6AI-4V) mini-implants of the same commercial brand, 1.5 mm in diameter, 8.0 mm in length and 2.0 mm transmucosal profile that were selected and divided into two groups Group 1 consisted of 24 mini-implants inserted in the maxilla and mandible of 17 patients under orthodontic treatment, and that after 4 months were removed and submitted to recycling processes, including cleaning and sterilization. Group 2 (control) consisted of 10 mini-implants as marketed by the manufacturer. The insertion torque values and the maximum torque at the time of the mini- implant fracture were performed on the universal testing machine (EMIC). To compare data between groups, independent t was performed (p> 0.05).

Results:

There was no statistically significant difference in the torsion resistance test between the groups analyzed.

Conclusion:

It can be concluded that the mini-implants recycling process did not influence values of torsion resistance, remaining within the limit recommended by the manufacturer for its clinical use. (AU)
Assuntos

Buscar no Google
Coleções: Bases de dados nacionais Contexto em Saúde: ODS3 - Saúde e Bem-Estar Problema de saúde: Meta 3.3: Acabar com as doenças tropicais negligenciadas e combater as doenças transmissíveis Base de dados: BBO - Odontologia Assunto principal: Ortodontia Corretiva / Implantes Dentários / Esterilização / Procedimentos de Ancoragem Ortodôntica Limite: Humanos Idioma: Português Revista: Ortho Sci., Orthod. sci. pract Ano de publicação: 2022 Tipo de documento: Artigo Instituição/País de afiliação: UFMG/BR / UNINGÁ/BR / USP/BR
Buscar no Google
Coleções: Bases de dados nacionais Contexto em Saúde: ODS3 - Saúde e Bem-Estar Problema de saúde: Meta 3.3: Acabar com as doenças tropicais negligenciadas e combater as doenças transmissíveis Base de dados: BBO - Odontologia Assunto principal: Ortodontia Corretiva / Implantes Dentários / Esterilização / Procedimentos de Ancoragem Ortodôntica Limite: Humanos Idioma: Português Revista: Ortho Sci., Orthod. sci. pract Ano de publicação: 2022 Tipo de documento: Artigo Instituição/País de afiliação: UFMG/BR / UNINGÁ/BR / USP/BR
...