Uso de parches protectores faciales no reduce la presión facial en un modelo simulado de ventilación mecánica no invasiva / Face protective patches do not reduce facial pressure ulcers in a simulated model of non-invasive ventilation
Riquelme M, Hugo; Wood V, David; Martínez F, Santiago; Carmona M, Fernando; Peña V, Axel; Wegner A, Adriana.
Afiliação
  • Riquelme M, Hugo; Complejo Asistencial Dr. Sótero del Río. Unidad de Paciente Crítico Pediátrico. Santiago. CL
  • Wood V, David; Complejo Asistencial Dr. Sótero del Río. Unidad de Paciente Crítico Pediátrico. Santiago. CL
  • Martínez F, Santiago; Complejo Asistencial Dr. Sótero del Río. Unidad de Paciente Crítico Pediátrico. Santiago. CL
  • Carmona M, Fernando; Complejo Asistencial Dr. Sótero del Río. Unidad de Paciente Crítico Pediátrico. Santiago. CL
  • Peña V, Axel; Complejo Asistencial Dr. Sótero del Río. Unidad de Paciente Crítico Pediátrico. Santiago. CL
  • Wegner A, Adriana; Complejo Asistencial Dr. Sótero del Río. Unidad de Paciente Crítico Pediátrico. Santiago. CL
Rev. chil. pediatr ; 88(3): 354-359, jun. 2017. ilus, tab
Artigo em Espanhol | LILACS | ID: biblio-899987
Biblioteca responsável: BR1.1
RESUMEN
La ventilación mecánica no invasiva (VMNI) tiene como complicación frecuente el desarrollo de úlceras faciales por presión (UPP). Su prevención considera el uso empírico de parches protectores entre piel y mascarilla, para disminuir la presión ejercida por ésta.

Objetivos:

Evaluar el efecto de los parches protectores sobre la presión ejercida por la mascarilla facial, y su impacto en los parámetros ventilatorios programados.

Método:

Modelo simulado de VMNI binivelada usando mascarilla facial total en fantoma con vía aérea fisiológica (ALS PRO+) en posición supina. Se midió la presión en frente, mentón y pómulos, usando 3 tipos de parches protectores de uso habitual versus un grupo control, utilizando sensores de presión (Interlinks Electronics®). Se evaluaron los valores obtenidos con el modelo de mascarilla-parches protectores en las variables programadas flujo máximo inspiratorio (FMI), volumen corriente espirado (Vte) y presión positiva inspiratoria (IPAP), con ventilador Trilogy 100, Respironics®. La programación y registro de las variables fue efectuada en 8 oportunidades en cada grupo por operadores independientes.

Resultados:

No se observó disminución de la presión facial con ninguno de los parches protectores respecto al grupo control. Moltoprén aumentó la presión facial en todos los puntos de apoyo (p < 0,001), aumentó fuga, disminuyó FMI, Vte e IPAP (p < 0,001). Parches de hidrocoloide aumentaron la presión facial sólo en pómulo izquierdo, aumentaron la fuga y disminuyeron FMI. Parches de poliuretano no generaron cambios en la presión facial ni en variables ventilatorias.

Conclusión:

El uso de parches protectores de moltoprén, hidrocoloide y poliuretano transparente no contribuyó a la disminución de la presión facial. Se observó un efecto deletéreo de los parches de moltoprén e hidrocoloide sobre la administración de variables ventilatorias, concluyendo que el no uso de los parches protectores permitió una mejor administración de los parámetros programados.
ABSTRACT
Noninvasive ventilation (NIV) frequently involves the development of facial pressure ulcers (FPU). Its prevention considers the empirical use of protective patches between skin and mask, in order to reduce the pressure exerted by it.

Objectives:

To evaluate the effect of protective patches on the pressure exerted by the facial mask, and its impact on the programmed ventilatory parameters.

Method:

Bilevel NIV simulated model using full face mask in phantom with a physiological airway (ALS PRO +) in supine position. Forehead, chin and cheekbones pressure were measured using 3 types of standard protective patches versus a control group using pressure sensors (Interlinks Electronics®). The values obtained with the protective patches-mask model were evaluated in the programmed variables maximum inspiratory flow (MIF)), expired tidal volume (Vte) and positive inspiratory pressure (IPAP), with Trilogy 100 ventilator, Respironics®. The programming and recording of the variables was carried out in 8 opportunities in each group by independent operators.

Results:

There was no decrease in facial pressure with any of the protective patches compared to the control group. Moltopren increased facial pressure at all support points (p < 0.001), increased leakage, it decreased MIF, Vte and IPAP (p < 0.001). Hydrocolloid patches increased facial pressure only in the left cheekbone, increased leakage and decreased MIF. Polyurethane patches did not produce changes in facial pressure or ventilatory variables.

Conclusion:

The use of protective patches of moltopren, hydrocolloid and polyurethane transparent did not contribute to the decrease of the facial pressure. A deleterious effect of the moltopren and hydrocolloid patches was observed on the administration of ventilatory variables, concluding that the non-use of the protective patches allowed a better administration of the programmed parameters.
Assuntos


Texto completo: Disponível Coleções: Bases de dados internacionais Base de dados: LILACS Assunto principal: Pressão / Bandagens / Lesão por Pressão / Pressão Positiva Contínua nas Vias Aéreas / Ventilação não Invasiva / Máscaras Tipo de estudo: Estudo de etiologia / Estudo de avaliação / Estudo prognóstico Limite: Humanos Idioma: Espanhol Revista: Rev. chil. pediatr Assunto da revista: Pediatria Ano de publicação: 2017 Tipo de documento: Artigo País de afiliação: Chile Instituição/País de afiliação: Complejo Asistencial Dr. Sótero del Río/CL

Texto completo: Disponível Coleções: Bases de dados internacionais Base de dados: LILACS Assunto principal: Pressão / Bandagens / Lesão por Pressão / Pressão Positiva Contínua nas Vias Aéreas / Ventilação não Invasiva / Máscaras Tipo de estudo: Estudo de etiologia / Estudo de avaliação / Estudo prognóstico Limite: Humanos Idioma: Espanhol Revista: Rev. chil. pediatr Assunto da revista: Pediatria Ano de publicação: 2017 Tipo de documento: Artigo País de afiliação: Chile Instituição/País de afiliação: Complejo Asistencial Dr. Sótero del Río/CL
...