Validación de cinco pulsioxímetros / Validation of five pulse oximeters
Sánchez-Pérez, EA.; Lozano-Nuevo, JJ.; Huerta-Ramírez, S.; Cerda-Téllez, F.; Mendoza-Portillo, E..
Afiliação
  • Sánchez-Pérez, EA.; UNAM. MX
  • Lozano-Nuevo, JJ.; UNAM. MX
  • Huerta-Ramírez, S.; SSCDMX. Ciudad de México. MX
  • Cerda-Téllez, F.; SSCDMX. Ciudad de México. MX
  • Mendoza-Portillo, E.; SSCDMX. Ciudad de México. MX
Med. interna Méx ; 33(6): 723-729, nov.-dic. 2017. tab, graf
Artigo em Espanhol | LILACS | ID: biblio-954908
Biblioteca responsável: MX1.1
RESUMEN
Resumen: ANTECEDENTES: la pulsioxímetria estima la saturación arterial de oxígeno mediante la absorción de un haz de luz infrarroja por la oxihemoglobina. OBJETIVO: determinar la validez y la confiabilidad interinstrumento de cinco pulsioxímetros comúnmente usados en la práctica clínica. MATERIAL Y MÉTODO: estudio transversal, analítico, realizado del 1 de enero de 2015 al 30 de octubre de 2016, en el que se incluyeron pacientes mayores de 18 años de edad, con línea arterial para toma de gasometría y se midió simultáneamente la saturación de oxígeno con cinco pulsioxímetros de uso habitual. Se determinaron medianas y porcentajes de los cinco pulsioxímetros entre sí mediante prueba de Friedman. RESULTADOS: se incluyeron 101 pacientes, 63 hombres y 38 mujeres. Se encontró similitud en las mediciones realizadas por cada pulsioxímetro con p = 0.08; todos los oxímetros se correlacionaron de manera positiva contra la prueba patrón de referencia (correlación de Pearson). Oxímetro 1: r = 0.90; oxímetro 2: r = 0.64; oxímetro 3: r = 0.57; oxímetro 4: r = 0.84 y oxímetro 5: r = 0.89; el área bajo la curva mostró oxímetro 1 (0.89), oxímetro 2 (0.88), oxímetro 3 (0.87), oxímetro 4 (0.83), oxímetro 5 (0.85) con p < 0.05. CONCLUSIONES: el pulsioxímetro número 1 muestra el mejor rendimiento comparado con la gasometría. Al comparar los oxímetros entre sí el rendimiento es igual.
ABSTRACT
Abstract: BACKGROUND: Pulse oximetry estimates the arterial oxygen saturation by absorption of an infrared light beam by oxyhemoglobin. OBJECTIVE: To determine the intraclass validity and reliability of five pulse oximeters commonly used in clinical practice. MATERIAL AND METHOD: A cross-sectional analytical study was done from January 1st 2015 to October 30 2016, including patients over 18 years old, with arterial line for blood gas sample and simultaneously oxygen saturation was measured with 5 pulse oximeters commonly used. Medians and percentages of the 5 pulse oximeters were determined by Friedman's test RESULTS: There were included 101 patients, 63 men and 38 women, with similarity in the measurements performed by each pulse oximeter with p = 0.08; all oximeters were positively correlated against the standard gold test (Pearson's correlation). Oximeter 1: r = 0.90; oximeter 2: r = 0.64; oximeter 3: r = 0.57; oximeter 4: r = 0.84 and oximeter 5: r = 0.89. The area under the curve showed oximeter 1 (0.89), oximeter 2 (0.88), oximeter 3 (0.87), oximeter 4 (0.83), oximeter 5 (0.85) with p < 0.05. CONCLUSIONS: The pulse oximeter number 1 shows a better performance when compared to the gasometry. When the oximeters are compared to each other the performance is the same.


Texto completo: Disponível Coleções: Bases de dados internacionais Base de dados: LILACS Tipo de estudo: Estudo observacional / Estudo de prevalência / Estudo prognóstico Idioma: Espanhol Revista: Med. interna Méx Assunto da revista: Medicina Interna Ano de publicação: 2017 Tipo de documento: Artigo País de afiliação: México Instituição/País de afiliação: SSCDMX/MX / UNAM/MX

Texto completo: Disponível Coleções: Bases de dados internacionais Base de dados: LILACS Tipo de estudo: Estudo observacional / Estudo de prevalência / Estudo prognóstico Idioma: Espanhol Revista: Med. interna Méx Assunto da revista: Medicina Interna Ano de publicação: 2017 Tipo de documento: Artigo País de afiliação: México Instituição/País de afiliação: SSCDMX/MX / UNAM/MX
...