Validación del método m-hidroxibifenilo para la cuantificación de ácido urónico en polisacáridos purificados de Streptococcus pneumoniae / Validation of the m-hydroxibiphenil method for the quantitation of uronic acid in purified Streptococcus pneumoniae polysaccharides
Pérez Calixto, Marylín; Suárez Pérez, Yania; González Rodríguez, Humberto; Achong García, Lay; Borrell Castillo, Yohana.
Afiliação
  • Pérez Calixto, Marylín; Instituto Finlay, Centro de Investigación‒Producción de Vacunas y Sueros. Instituto Finlay. La Habana. Cuba
  • Suárez Pérez, Yania; Instituto de Farmacia y Alimentos (IFAL). Universidad de La Habana. La Habana. Cuba
  • González Rodríguez, Humberto; Instituto Finlay, Centro de Investigación‒Producción de Vacunas y Sueros. La Habana. Cuba
  • Achong García, Lay; Instituto Finlay, Centro de Investigación‒Producción de Vacunas y Sueros. La Habana. Cuba
  • Borrell Castillo, Yohana; Instituto Finlay, Centro de Investigación‒Producción de Vacunas y Sueros. Instituto Finlay. La Habana. Cuba
Rev. cuba. farm ; 49(2)abr.-jun. 2015. graf, tab
Artigo em Espanhol | CUMED | ID: cum-63292
Biblioteca responsável: CU1.1
Localização: CU1.1
RESUMEN

Introducción:

la complejidad y diversidad estructural de los polisacáridos capsulares de Streptococcus pneumoniae hace que la metodología analítica necesaria para la caracterización y control de calidad sea engorrosa. El método colorimétrico m‒hidroxibifenil desarrollado por Blumenkrantz y Asboe‒Hansen en el año 1973, permite contrarrestar las desventajas del método propuesto en la Farmacopea Europea 6ta Edición, 2011.

Objetivo:

validar el método m‒hidroxibifenilo para cuantificar ácido urónico en polisacáridos purificados de streptococcus pneumoniae.

Métodos:

se determinó el tiempo de vigencia de la solución de tetraborato de sodio responsable de generar la respuesta analítica y se seleccionó una metodología para la cuantificación de ácido urónico por el método del m‒hidroxibifenilo se emplea una solución estándar de ácido galacturónico a 1 mg/mL. Se realizó una evaluación de los parámetros de validación linealidad, exactitud, precisión, rango y especificidad, según exigencias actuales. Además se comparó con el método del carbazol descrito en la Farmacopea Europea 6ta Edición, 2011 para la cuantificación de ácido urónico.

Resultados:

se demostró que el tiempo de vida útil de la solución de tetraborato de sodio fue de 24 h. El método del m‒hidroxibifenilo fue específico, lineal, exacto y preciso en el rango de 2‒20 μg/mL porque se cumplieron satisfactoriamente los criterios de aceptación establecidos para cada uno de ellos. La comparación del método propuesto con el método normalizado reveló que no existieron diferencias estadísticamente significativas entre ambos.

Conclusión:

el método colorimétrico del m‒hidroxibifenilo resultó válido para el control de calidad de las muestras de polisacárido purificado de Streptococcus pneumoniae, dando resultados comparables con el método recomendado en la Farmacopea Europea, 2011(AU)
ABSTRACT

Introduction:

complexity and diversity in the structure of Streptococcus pneumoniae capsular makes the analytical methodology for characterization and quality control a troublesome aspect. The colorimetric m/hydroxibiphenil method devised by Blumenklrantz and Asboe-Hansen in 1973 allows reducing the disadvantages of the suggested method in the European Pharmacopea 6th edition, 2011.

Objective:

to validate the m-hydroxybiphenil method for quantitation of uronic acid in purified Streptococcus pneumonia polysaccharides.

Methods:

the length of validity of a sodium tetraborate solution, responsible for generating the analytical response, was estimated, and additionally, a methodology for quantitation of uronic acid by the m-hydroxybiphenil method was selected in which a standard galacturonic acid solution at 1mg/ml was used. The validation parameters were evaluated as follows linearity, accuracy, precision, range and specificity, according to the present requirements. This method was compared with the carbazol method described in the European Pharmacopea 6th edition, 2011 for quantitation of uronic acid.

Results:

it was demonstrated that the useful lifetime of the sodium tetraborate solution was 24 hours. The m-hydroxybiphenil method was specific, linear, accurate and precise in the range of 2 to 20 ?g/mL because the set acceptance criteria were satisfactorily complied with for each of them. The comparison of the suggested method with the standardized method yielded that no statistically significant differences exist between them.

Conclusions:

the colorimetric m-hydroxybiphenil method proved to be valid for the quality control of purified Streptococcus pneumonia polysaccharide samples and the results are comparable to those of the recommended method in the European Pharmacopea, 2011(AU)
Assuntos

Texto completo: Disponível Coleções: Bases de dados nacionais / Cuba Contexto em Saúde: Doenças Negligenciadas Problema de saúde: Hanseníase Base de dados: CUMED Assunto principal: Infecções Pneumocócicas Idioma: Espanhol Revista: Rev. cuba. farm Ano de publicação: 2015 Tipo de documento: Artigo Instituição/País de afiliação: Instituto Finlay, Centro de Investigación‒Producción de Vacunas y Sueros/Cuba / Instituto de Farmacia y Alimentos (IFAL). Universidad de La Habana/Cuba
Texto completo: Disponível Coleções: Bases de dados nacionais / Cuba Contexto em Saúde: Doenças Negligenciadas Problema de saúde: Hanseníase Base de dados: CUMED Assunto principal: Infecções Pneumocócicas Idioma: Espanhol Revista: Rev. cuba. farm Ano de publicação: 2015 Tipo de documento: Artigo Instituição/País de afiliação: Instituto Finlay, Centro de Investigación‒Producción de Vacunas y Sueros/Cuba / Instituto de Farmacia y Alimentos (IFAL). Universidad de La Habana/Cuba
...