Determinación de óxido nítrico en aire espirado (FENO) mediante un equipo portátil (NIOX-MINO® Aerocrine) en población sana / Measurement of fraction of exhaled nitric oxide with the portable NIOX-MINO monitor in healthy adults
Fortuna, Ana María; Feixas, Teresa; Casan, Pere.
Afiliação
  • Fortuna, Ana María; Universidad Autónoma de Barcelona. Hospital de la Santa Creu i Sant Pau. Barcelona. España
  • Feixas, Teresa; Universidad Autónoma de Barcelona. Hospital de la Santa Creu i Sant Pau. Barcelona. España
  • Casan, Pere; Universidad Autónoma de Barcelona. Hospital de la Santa Creu i Sant Pau. Barcelona. España
Arch. bronconeumol. (Ed. impr.) ; 43(3): 176-179, mar. 2007. ilus, tab
Artigo em Es | IBECS | ID: ibc-052290
Biblioteca responsável: ES1.1
Localização: ES1.1 - BNCS
RESUMEN
La determinación de óxido nítrico en aire espirado (FENO) es una técnica no invasiva que se utiliza en la evaluación clínica y el tratamiento del asma. El objetivo de este estudio ha sido determinar los valores de referencia de FENO en un grupo de voluntarios sanos mediante la utilización de un nuevo equipo portátil (NIOX-MINO® Aerocrine), así como determinar el grado de relación que presenta con el equipo de determinación habitual en nuestro laboratorio de función pulmonar (sensor de quimioluminiscencia N-6008® SIR). Según los resultados obtenidos, los valores de FENO que da el equipo portátil son siempre superiores a los que ofrece el sensor habitual, con un valor de corte de 34 ppb (media + 2 desviación estándar). Existe un relación directa y significativa de la determinación de FENO entre ambos equipos (r = 0,92; p = 0,001) con un factor de corrección de FENO (NIOX-MINO®) = 10 + 1,5 FENO (N-6008®). La relación entre los valores de FENO y la edad, el sexo, el índice de masa corporal y los valores espirométricos no fue estadísticamente significativa
ABSTRACT
Measurement of the fraction of exhaled nitric oxide (FENO) provides a noninvasive way to monitor asthma treatment in clinical practice. The aim of this study was to determine FENO reference values for measurements recorded with the portable NIOX MINO monitor in a group of healthy volunteers. We also assessed the association between values recorded by the portable monitor and the N-6008 chemiluminescence analyzer used in our pulmonary function laboratory. The FENO values obtained with the portable monitor were consistently higher than those recorded by the N-6008 analyzer; the cutoff value for the portable monitor was 34 ppb (mean + 2 SD). We detected a direct correlation (r=0.92) between the FENO measurements recorded by the 2 monitors (P=.001). The following equation expresses the relationship between measurements from the 2 devices FENO(NIOX MINO) = 10 + [1.5 FENO(N-6008)]. We did not observe statistically significant correlations between FENO measurements and age, sex, body mass index, or spirometry
Assuntos
Buscar no Google
Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Asma / Expiração / Óxido Nítrico Limite: Humanos Idioma: Espanhol Revista: Arch. bronconeumol. (Ed. impr.) Ano de publicação: 2007 Tipo de documento: Artigo Instituição/País de afiliação: Universidad Autónoma de Barcelona/España
Buscar no Google
Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Asma / Expiração / Óxido Nítrico Limite: Humanos Idioma: Espanhol Revista: Arch. bronconeumol. (Ed. impr.) Ano de publicação: 2007 Tipo de documento: Artigo Instituição/País de afiliação: Universidad Autónoma de Barcelona/España
...