Your browser doesn't support javascript.
loading
Stents traqueales metálicos autoexpandibles. Estudio comparativo de 3 tipos diferentes de stents en un modelo animal / Tracheal Self-Expandable Metallic Stents: A Comparative Study of Three Different Stents in a Rabbit Model
Serrano, Carolina; Lostalé, Fernando; Rodríguez-Panadero, Francisco; Blas, Ignacio de; Laborda, Alicia; Gregorio, Miguel Angel de.
Afiliação
  • Serrano, Carolina; Universidad de Zaragoza. Zaragoza. España
  • Lostalé, Fernando; Universidad de Zaragoza. Zaragoza. España
  • Rodríguez-Panadero, Francisco; Hospital Virgen del Rocío/CSIC/Universidad de Sevilla. Instituto de Biomedicina de Sevilla (IBIS). Sevilla. España
  • Blas, Ignacio de; Universidad de Zaragoza. Departamento de Patología Animal. Zaragoza. España
  • Laborda, Alicia; Universidad de Zaragoza. Zaragoza. España
  • Gregorio, Miguel Angel de; Universidad de Zaragoza. Zaragoza. España
Arch. bronconeumol. (Ed. impr.) ; 52(3): 123-130, mar. 2016. ilus, tab
Artigo em Espanhol | IBECS | ID: ibc-149910
Biblioteca responsável: ES1.1
Localização: BNCS
RESUMEN

Introducción:

El objetivo de este estudio es evaluar l8mm frea reactividad traqueal tras la implantación de distintos stents metálicos autoexpandibles (SMAE). Material y

métodos:

Se utilizaron 40 conejos hembra de raza neozelandesa, que se dividieron en 4 grupos. En 3 grupos se implantaron SMAE de acero (SA), de nitinol (NiTi) o stents liberadores de nitinol (SLF). El cuarto grupo fue el grupo de control (sin stent). Los stents se implantaron por vía percutánea bajo control fluoroscópico. Los animales se evaluaron mediante tomografía axial computarizada (TAC) multicorte y las tráqueas se extirparon para su estudio anatomopatológico (EAP). Los datos de la TAC y el EAP se analizaron estadísticamente y se correlacionaron.

Resultados:

El grupo que recibió SLF presentaba la mayor longitud de estenosis (20,51 ± 14,0nte a 5,84 ± 12,43 y 6,57 ± 6,54 mm en los grupos NiTi y SA, día 30; p < 0,05) y el mayor índice de formación de granulomas evidenciados mediante TAC (50% de los casos). El grupo al que se implantaron stents NiTi mostró el menor grado de estenosis (2,86 ± 6,91% frente a 11,28 ± 13,98 y 15,54 ± 25,95% en los grupos SLF y SA; p<0,05). En el estudio AP, el grupo SA presentó reactividad proliferativa intensa en comparación con los otros 2 grupos. En el grupo SLF se observó una respuesta destructiva en el 70% de animales, mientras que el stent NiTi fue el que menos reacción provocó. La TAC resultó ser superior para detectar el engrosamiento (correlación positiva de un 68,9%; p < 0,001) que para la observación de granulomas (n.s.).

Conclusiones:

El grupo SA desarrolló granulomas y estenosis significativas. El stent NiTi fue el que menos reacción indujo, mientras que el SLN provocó lesiones importantes que podrían estar relacionadas con la dosis de fármaco. Por consiguiente, este tipo de SLF no se recomienda para el tratamiento de la estenosis traqueobronquial
ABSTRACT

Introduction:

The objective of this study was to assess tracheal reactivity after the deployment of different self-expandable metal stents (SEMS). Material and

methods:

Forty female New Zealand rabbits were divided into four groups. Three groups received three different SEMS steel (ST), nitinol (NiTi), or nitinol drug-eluting stent (DES); the fourth group was the control group (no stent). Stents were deployed percutaneously under fluoroscopic guidance. Animals were assessed by multi-slice, computed tomography (CT) scans, and tracheas were collected for anatomical pathology (AP) study. Data from CT and AP were statistically analyzed and correlated.

Results:

The DES group had the longest stenosis (20.51 ± 14.08 mm vs. 5.84 ± 12.43 and 6.57 ± 6.54 mm in NiTi and ST, respectively, day 30; P < .05), and higher granuloma formation on CT (50% of cases). The NiTi group showed the lowest grade of stenosis (2.86 ± 6.91% vs. 11.28 ± 13.98 and 15.54 ± 25.95% in DES and ST, respectively; P<.05). The AP study revealed that the ST group developed intense proliferative reactivity compared to the other groups. In the DES group, a destructive response was observed in 70% of the animals, while the NiTi was the least reactive stent. CT was more effective in detecting wall thickening (positive correlation of 68.9%; P < .001) than granuloma (not significant).

Conclusions:

The ST group developed granulomas and significant stenosis. NiTi was the least reactive stent, while DES caused significant lesions that may be related to drug dosage. This type of DES stent is therefore not recommended for the treatment of tracheobronchial stenosis
Assuntos

Texto completo: Disponível Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Estenose Traqueal / Obstrução das Vias Respiratórias / Stents Farmacológicos Tipo de estudo: Guia de prática clínica / Estudo prognóstico Limite: Animais Idioma: Espanhol Revista: Arch. bronconeumol. (Ed. impr.) Ano de publicação: 2016 Tipo de documento: Artigo Instituição/País de afiliação: CSIC / Universidad de Zaragoza/España

Texto completo: Disponível Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Estenose Traqueal / Obstrução das Vias Respiratórias / Stents Farmacológicos Tipo de estudo: Guia de prática clínica / Estudo prognóstico Limite: Animais Idioma: Espanhol Revista: Arch. bronconeumol. (Ed. impr.) Ano de publicação: 2016 Tipo de documento: Artigo Instituição/País de afiliação: CSIC / Universidad de Zaragoza/España
...