Evaluación de un método para la medida de la concentración de ferritina en un analizador RxL© / Evaluation of a method for the measurement of ferritin concentration in an Analyzer RxL©
Baz, MJ; Bueno MJ; Castro, M; Carneiro, I.
Afiliação
  • Baz, MJ; Complejo Hospitalario Llerena-Zafra. Badajoz. España
  • Bueno MJ; Complejo Hospitalario Llerena-Zafra. Badajoz. España
  • Castro, M; Complejo Hospitalario Llerena-Zafra. Badajoz. España
  • Carneiro, I; Complejo Hospitalario Llerena-Zafra. Badajoz. España
Rev. diagn. biol ; 53(4): 180-184, oct.-dic. 2004. tab, graf
Artigo em Espanhol | IBECS | ID: ibc-157105
Biblioteca responsável: ES1.1
Localização: BNCS
RESUMEN
El objetivo de este trabajo ha sido evaluar la calidad analítica de la determinación dela concentración de ferritina mediante un inmunoensayo heterogéneo en el analizador Dimensión' RxL. Se ha evaluado la imprecisión intraserial y la interserial, la linealidad, la recuperación, el límite de detección, la contaminación por arrastre entre especimenes, la exactitud, la interferencia por bilirrubina, hemólisis y triglicéridos, asi como la comparación con otro inmunoensayo. Los coeficientes de variación intra e interserial indicaron una buena reproducibilidad, en todos los casos fue inferior al 6,5%. El procedimiento evaluado ha sido lineal dentro del intervalo de medida. La recuperación estuvo próxima al 100%. No existió contaminación por arrastre entre especimenes y no presentó interferencias significas por bilirrubina, hemólisis y triglicéridos. Se observó buena correlación entre los resultados obtenidos por los dos métodos comparados, aunque no fueron intercambiables. En conclusión, puede considerarse un procedimiento adecuado para la medida rápida, fiable y precisa de ferritina en suero (AU)
ABSTRACT
The aim of this study was to evaluate the analytical performance of the measurement of ferritin concentration in serum, using a heterogeneus immunoassay method on the Dimensión' RxL analyser. We have evaluated the repetibility intra and interserial, the linearty, recovery, carry-over between samples, detection limit, bilirrubin, hemglobin and triglycerides interference and it was performed comparison with other immnoassay method. Either the intra and interserial repetibily comply the acceptability criteria The essay is linear over the calibration range. Recovery was closed to 100%. On the other hand, it was cheked thas does not exit contamination effect between samples. No significatives interference occurred from bilirrubin, hemoglobin and trigiycerides We obtained good correlation between the values obtanined by both methods. In conclusion, we consider this procedure as a reliable method for serum ferritin measurement (AU)
Assuntos
Buscar no Google
Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Bioensaio / Imunoensaio / Compostos de Cromo / Soro / Ferritinas Tipo de estudo: Estudo diagnóstico / Estudo de avaliação Limite: Feminino / Humanos / Masculino Idioma: Espanhol Revista: Rev. diagn. biol Ano de publicação: 2004 Tipo de documento: Artigo Instituição/País de afiliação: Complejo Hospitalario Llerena-Zafra/España
Buscar no Google
Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Bioensaio / Imunoensaio / Compostos de Cromo / Soro / Ferritinas Tipo de estudo: Estudo diagnóstico / Estudo de avaliação Limite: Feminino / Humanos / Masculino Idioma: Espanhol Revista: Rev. diagn. biol Ano de publicação: 2004 Tipo de documento: Artigo Instituição/País de afiliação: Complejo Hospitalario Llerena-Zafra/España
...