Proyecto Girona Territori Cardioprotegit: evaluación del funcionamiento de los desfibriladores públicos / The Girona Territori Cardioprotegit project: performance evaluation of public defibrillators
Loma-Osorio, Pablo; Nuñez, Maria; Aboal, Jaime; Bosch, Daniel; Batlle, Pau; Ruiz de Morales, Ester; Ramos, Rafael; Brugada, Josep; Onaga, Hisao; Morales, Alex; Olivet, Josep; Brugada, Ramon.
Afiliação
  • Loma-Osorio, Pablo; Universidad de Girona. Facultad de Medicina. Departamento de Ciencias Médicas. Girona. España
  • Nuñez, Maria; Hospital Universitari Dr Josep Trueta. Servicio de Cardiología. Girona. España
  • Aboal, Jaime; Hospital Universitari Dr Josep Trueta. Servicio de Cardiología. Girona. España
  • Bosch, Daniel; Hospital Universitari Dr Josep Trueta. Servicio de Cardiología. Girona. España
  • Batlle, Pau; Diputación de Girona (Dipsalut). Departamento de Salud. Girona. España
  • Ruiz de Morales, Ester; Diputación de Girona (Dipsalut). Departamento de Salud. Girona. España
  • Ramos, Rafael; Instituto de Investigación en Atención Primaria (IDIAP Jordi Gol). Girona. España
  • Brugada, Josep; Hospital Clínic i Provincial de Barcelona. Instituto del Tórax. Barcelona. España
  • Onaga, Hisao; Región Sanitaria de Girona. Servicio de Emergencias Médicas (SEM). Girona. España
  • Morales, Alex; Diputación de Girona (Dipsalut). Departamento de Salud. Girona. España
  • Olivet, Josep; Universidad de Girona. Grupo de Investigación, Salud y Atención Sanitaria. Girona. España
  • Brugada, Ramon; Instituto de investigación biomédica de Girona (IDIBGI). Girona. España
Rev. esp. cardiol. (Ed. impr.) ; 71(2): 79-85, feb. 2018. graf, tab, ilus
Artigo em Espanhol | IBECS | ID: ibc-170657
Biblioteca responsável: ES1.1
Localização: BNCS
RESUMEN
Introducción y

objetivos:

La desfibrilación pública ha potenciado la difusión y el uso de los desfibriladores externos automáticos (DEA) de manera exponencial, pero se carece de datos recientes sobre su uso en nuestro medio. Se realizó un análisis descriptivo del funcionamiento de los DEA desde el despliegue de un programa de desfibrilación pública.

Métodos:

Análisis retrospectivo de los trazados electrocardiográficos registrados y del funcionamiento de los desfibriladores, en el marco de un programa de desfibrilación pública en la provincia de Girona, desde su inicio en junio de 2011 hasta junio de 2015.

Resultados:

De 231 activaciones, se dispuso de información completa en 188, entre dispositivos móviles (82%) y fijos (18%). La asistolia fue el ritmo más prevalente (42%), mientras que un 23% de los ritmos fueron fibrilación ventricular. La especificidad de los dispositivos identificando ritmos desfibrilables fue del 100%, pero hubo 8 falsos negativos (sensibilidad del 83%). En el caso de los 47 ritmos desfibrilables, la tasa de recuperación de la circulación espontánea fue del 49% (23 casos). No hubo ningún accidente relacionado con el uso del dispositivo.

Conclusiones:

Casi la mitad de los ritmos registrados fueron asistolias. Los DEA analizados presentaron excelentes seguridad y especificidad, con una sensibilidad moderada. El DEA trató con éxito a la mitad de los pacientes que presentaron ritmo desfibrilable (AU)
ABSTRACT
Introduction and

objectives:

In recent years, public access defibrillation programs have exponentially increased the availability of automatic external defibrillators (AED) in public spaces but there are no data on their performance in our setting. We conducted a descriptive analysis of the performance of AED since the launch of a public defibrillation program in our region.

Methods:

A retrospective analysis was conducted of electrocardiographic tracings and the performance of AED in a public defibrillation program from June 2011 to June 2015 in the province of Girona, Spain.

Results:

There were 231 AED activations. Full information was available on 188 activations, of which 82% corresponded to mobile devices and 18% to permanent devices. Asystole was the most prevalent rhythm (42%), while ventricular fibrillation accounted for 23%. The specificity of the device in identifying a shockable rhythm was 100%, but there were 8 false negatives (sensitivity 83%). There were 47 shockable rhythms, with a spontaneous circulation recovery rate of 49% (23 cases). There were no accidents related to the use of the device (AU)
Assuntos

Texto completo: Disponível Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Fibrilação Ventricular / Sensibilidade e Especificidade / Reanimação Cardiopulmonar / Desfibriladores / Projetos / Parada Cardíaca Tipo de estudo: Estudo diagnóstico / Estudo observacional Limite: Humanos Idioma: Espanhol Revista: Rev. esp. cardiol. (Ed. impr.) Ano de publicação: 2018 Tipo de documento: Artigo Instituição/País de afiliação: Diputación de Girona (Dipsalut)/España / Hospital Clínic i Provincial de Barcelona/España / Hospital Universitari Dr Josep Trueta/España / Instituto de Investigación en Atención Primaria (IDIAP Jordi Gol)/España / Instituto de investigación biomédica de Girona (IDIBGI)/España / Región Sanitaria de Girona/España / Universidad de Girona/España

Texto completo: Disponível Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Fibrilação Ventricular / Sensibilidade e Especificidade / Reanimação Cardiopulmonar / Desfibriladores / Projetos / Parada Cardíaca Tipo de estudo: Estudo diagnóstico / Estudo observacional Limite: Humanos Idioma: Espanhol Revista: Rev. esp. cardiol. (Ed. impr.) Ano de publicação: 2018 Tipo de documento: Artigo Instituição/País de afiliação: Diputación de Girona (Dipsalut)/España / Hospital Clínic i Provincial de Barcelona/España / Hospital Universitari Dr Josep Trueta/España / Instituto de Investigación en Atención Primaria (IDIAP Jordi Gol)/España / Instituto de investigación biomédica de Girona (IDIBGI)/España / Región Sanitaria de Girona/España / Universidad de Girona/España
...