Fiabilidad interexaminador de la exploración de puntos gatillo miofasciales en la musculatura de las regiones cervical y lumbar / Inter-examiner reliability of the examination of trigger points in the cervical and lumbar spine regions
Poveda-Pagán, EJ; Catalán García, I; Santa Lozano, A; Segura Heras, JV; Lozano-Quijada, C.
Afiliação
  • Poveda-Pagán, EJ; Universidad Miguel Hernández. Centro de Investigación Traslacional en Fisioterapia. Área de Fisioterapia, Departamento de Patología y Cirugía. Elche. España
  • Catalán García, I; s.af
  • Santa Lozano, A; s.af
  • Segura Heras, JV; Universidad Miguel Hernández. Centro de Investigación Operativa. Elche. España
  • Lozano-Quijada, C; Universidad Miguel Hernández. Centro de Investigación Traslacional en Fisioterapia. Área de Fisioterapia, Departamento de Patología y Cirugía. Elche. España
Fisioterapia (Madr., Ed. impr.) ; 40(2): 79-87, mar.-abr. 2018. ilus, graf, tab
Artigo em Espanhol | IBECS | ID: ibc-171700
Biblioteca responsável: ES1.1
Localização: BNCS
RESUMEN

Objetivo:

Conocer el grado de concordancia y la correlación interexaminadores al realizar la exploración manual de las características de banda tensa, banda sensible, dolor referido, dolor familiar, la respuesta de espasmo local y la algometría en los puntos gatillo explorados. Material y

método:

Estudio observacional prospectivo con 30 sujetos, 8 hombres (media 36,55 años; desviación estándar7,21) y 22 mujeres (44,53 años; desviación estándar 0,81). Los examinadores fueron 2 fisioterapeutas noveles (N1 y N2) y un examinador experto (E1). Los músculos explorados esternocleidomastoideo, trapecio superior, elevador de escápula, infraespinoso, cuadrado lumbar, glúteo medio y piramidal.

Resultados:

Kappa sustancial o casi perfecto entre examinadores para el lado dominante, excepto el dolor referido en el músculo piramidal con 0,35 (intervalo confianza -IC-95 por mil 0 a 0,92) entre N1 frente a N2, 0,20(0 a 0,66) entre E1 frente a N2 y 0,27(0 a 0,77) entre E1 frente a N1 y la respuesta de espasmo local con 0,34(0 a 0,70) entre N1 frente a N2 para el músculo infraespinoso y 0(0,13 a 0,80) para el músculo trapecio superior. La correlación intraclase ha sido buena o muy buena para todos los músculos explorados, exceptuando el músculo elevador de escápula entre E1 frente a N1 con 0,68(0,44 a 0,83) y entre E1 frente a N2 con 0,69(0,45 a 0,84).

Conclusiones:

Obtenemos valor Kappa sustancial y/o casi perfecto entre examinadores, exceptuando en el dolor referido y la respuesta de espasmo local en algún músculo. Por otro lado, en el CCI para los valores de algometría, muestran una correlación buena o muy buena para todos los músculos a excepción del elevador de escápula
ABSTRACT

Objective:

To determine the level of agreement and inter-rater reliability when performing the manual examination of the characteristics of taut band, sensitive edge, referred pain, familiar pain, local twitch response, and the algometry explored.

Methods:

A prospective observational study was conducted on 30 subjects including 8 men (mean 36.55 years, standard deviation 7.21) and 22 women (44.53 years old, standard deviation 0.81). The examiners were 2 non-expert physiotherapists (N1 and N2) and an expert examiner (E1). Muscles explored sternocleidomastoid, upper trapezius, levator scapula, infraspinatus, lumbar square, gluteus medius, and piriform.

Results:

Kappa was substantial or almost perfect between examiners for the dominant side, except pain referred to in the piriform muscle with 0.35 (confidence interval -95 per-mille CI 0 - 0.92) between N1 vs. N2, 0.20 (0 - 0.66) between E1 vs. N2 and 0.27 (0 - 0.77) between E1 vs. N1 and local twitch response with 0.34 (0 - 0.70) between N1 vs. N2 for the infraspinatus muscle and 0 (0.13 - 0,80) for the upper trapezius muscle. Intraclass correlation was good or very good for all explored muscles, except for the levator scapula muscle between E1 vs. N1 with 0.68 (0.44 to 0.83) and between E1 vs. N2 with 0.69 (0.45 to 0.84).

Conclusions:

A substantial and/or almost perfect agreement (Kappa value) was obtained between examiners, except in the referred pain and local twitch response in some muscles. On the other hand, intraclass correlation for algometry values showed a good or very good correlation for all muscles except the levator scapula
Assuntos
Buscar no Google
Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Limiar da Dor / Dor nas Costas / Lesões do Pescoço / Pontos-Gatilho / Fisioterapeutas / Músculos do Pescoço Tipo de estudo: Estudo observacional Limite: Adulto / Feminino / Humanos / Masculino Idioma: Espanhol Revista: Fisioterapia (Madr., Ed. impr.) Ano de publicação: 2018 Tipo de documento: Artigo Instituição/País de afiliação: Universidad Miguel Hernández/España
Buscar no Google
Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Limiar da Dor / Dor nas Costas / Lesões do Pescoço / Pontos-Gatilho / Fisioterapeutas / Músculos do Pescoço Tipo de estudo: Estudo observacional Limite: Adulto / Feminino / Humanos / Masculino Idioma: Espanhol Revista: Fisioterapia (Madr., Ed. impr.) Ano de publicação: 2018 Tipo de documento: Artigo Instituição/País de afiliação: Universidad Miguel Hernández/España
...