Your browser doesn't support javascript.
loading
Visual and satisfaction results with implantation of the trifocal Panoptix (R) intraocular lens in cataract surgery
Galvis, Virgilio; Escaf, Luis C; Escaf, Luis J; Tello, Alejandro; Rodríguez, Lisi D; Lapid-Gortzak, Ruth; Carreño, Néstor I; Berrospi, Rubén D; Niño, Camilo A; Viberg, Andreas; Camacho, Paul A.
Afiliação
  • Galvis, Virgilio; Centro Oftalmológico Virgilio Galvis. Fundación Oftalmológica de Santander FOSCAL. Universidad Autónoma de Bucaramanga UNAB. Colombia
  • Escaf, Luis C; Fundación Oftalmológica de Santander FOSCAL. Universidad Autónoma de Bucaramanga UNAB. Colombia
  • Escaf, Luis J; Clínica de Oftalmología del Caribe COFCA. Barranquilla. Colombia
  • Tello, Alejandro; Centro Oftalmológico Virgilio Galvis. Fundación Oftalmológica de Santander FOSCAL. Universidad Autónoma de Bucaramanga UNAB. Colombia
  • Rodríguez, Lisi D; Centro Oftalmológico Virgilio Galvis. Universidad Autónoma de Bucaramanga UNAB. Colombia
  • Lapid-Gortzak, Ruth; Amsterdam Universitary Medical Centers. Amsterdam. The Netherlands
  • Carreño, Néstor I; Centro Oftalmológico Virgilio Galvis. Colombia
  • Berrospi, Rubén D; Centro Oftalmológico Virgilio Galvis. Colombia
  • Niño, Camilo A; Centro Oftalmológico Virgilio Galvis. Colombia
  • Viberg, Andreas; Umeå University. Department of Clinical Science /Ophthalmology. Umeå. Sweden
  • Camacho, Paul A; Centro Oftalmológico Virgilio Galvis. Colombia
J. optom. (Internet) ; 15(3): 1-9, Jul-Sep.2022. tab, graf
Artigo em Inglês | IBECS | ID: ibc-204706
Biblioteca responsável: ES1.1
Localização: ES15.1 - BNCS
ABSTRACT

Purpose:

To assess visual results (including the Lapid-Kushlin defocus coefficient), and satisfaction, following the implantation of PanOptix® a trifocal intraocular lens.

Methods:

Observational, retrospective cohort study. 130 eyes of 65 subjects with bilateral implantation were included. Binocular uncorrected visual acuities [distance (Binocular UDVA), intermediate (Binocular UIVA) and near (Binocular UNVA)] were measured. Also, the manifest refraction, binocular defocus curve and binocular contrast sensitivity were determined. The Lapid-Kushlin defocus coefficient, a new single numerical parameter based on the area under the curve of the defocus curve, was calculated. The validated Spanish CATQUEST-9SF satisfaction survey was applied.

Results:

Postoperative visual acuity ranges were binocular UDVA 0.18 to 0.0 LogMAR, binocular UIVA 0.30 to 0.0 LogMAR and Binocular UNVA 0.18 to 0.0 LogMAR. The levels of contrast sensitivity both in bright light conditions (with and without glare) and low light conditions with glare, remained within the limits of normality. Under scotopic conditions without glare values below normality were found at lower spatial frequencies. The binocular defocus curve showed a plateau without a clearly evident peak. The Lapid-Kushlin defocus coefficient was 0.199. Two patients (3.1%) needed glasses after the procedure for near and intermediate vision. No patient manifested great visual difficulties or was dissatisfied with the results.

Conclusions:

The trifocal platform showed very good results in this series of patients. 96.9% of the patients achieved independence of the glasses and expressed a high degree of satisfaction. The Lapid-Kushlin defocus coefficient was better than those calculated from literature, for other multifocal intraocular lenses. (AU)
Assuntos


Texto completo: Disponível Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Satisfação Pessoal / Desenho de Prótese / Refração Ocular / Catarata / Satisfação do Paciente / Lentes Intraoculares Multifocais Limite: Humanos Idioma: Inglês Revista: J. optom. (Internet) Ano de publicação: 2022 Tipo de documento: Artigo Instituição/País de afiliação: Amsterdam Universitary Medical Centers/The Netherlands / Centro Oftalmológico Virgilio Galvis/Colombia / Clínica de Oftalmología del Caribe COFCA/Colombia / Fundación Oftalmológica de Santander FOSCAL/Colombia / Umeå University/Sweden

Texto completo: Disponível Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Satisfação Pessoal / Desenho de Prótese / Refração Ocular / Catarata / Satisfação do Paciente / Lentes Intraoculares Multifocais Limite: Humanos Idioma: Inglês Revista: J. optom. (Internet) Ano de publicação: 2022 Tipo de documento: Artigo Instituição/País de afiliação: Amsterdam Universitary Medical Centers/The Netherlands / Centro Oftalmológico Virgilio Galvis/Colombia / Clínica de Oftalmología del Caribe COFCA/Colombia / Fundación Oftalmológica de Santander FOSCAL/Colombia / Umeå University/Sweden
...