Evaluación de la idoneidad y aplicabilidad clínica de diferentes electrodos para la monitorización aEEG/cEEG en el niño prematuro extremo / Evaluation of the suitability and clinical applicability of different electrodes for aEEG/cEEG monitoring in the extremely premature infant
Cordeiro, Malaika; Peinado, Helena; Teresa Montes, María; Valverde, Eva.
Afiliação
  • Cordeiro, Malaika; Hospital Universitario La Paz. Servicio de Neonatología. Fundación NeNe. Madrid. España
  • Peinado, Helena; Hospital Universitario La Paz. Servicio de Neonatología. Madrid. España
  • Teresa Montes, María; Hospital Universitario La Paz. Servicio de Neonatología. Fundación NeNe. Madrid. España
  • Valverde, Eva; Hospital Universitario La Paz. Servicio de Neonatología. Fundación NeNe. Madrid. España
An. pediatr. (2003. Ed. impr.) ; 95(6): 423-430, Dic. 2021. ilus
Artigo em Espanhol | IBECS | ID: ibc-208365
Biblioteca responsável: ES1.1
Localização: ES15.1 - BNCS
RESUMEN

Introducción:

La neuromonitorización de la función cerebral mediante electroencefalografía continua (aEEG/cEEG) es una herramienta esencial en el cuidado estándar del niño a término, y de utilidad creciente en el niño prematuro como biomarcador de lesión y maduración cerebral. Sin embargo, la colocación de los electrodos supone un gran reto, especialmente en el niño prematuro extremo, desalentando frecuentemente su neuromonitorización. El objetivo de este estudio es analizar los diferentes electrodos disponibles, seleccionar el que mejor se adapta a las peculiaridades del niño prematuro extremo y evaluar su aplicabilidad en la práctica clínica.Población y

métodos:

Con motivo del diseño de un protocolo de estudio de neuromonitorización mediante aEEG/cEEG en niños prematuros<28 semanas, analizamos nuestra experiencia con los tipos de electrodos disponibles y seleccionamos el que consideramos más adecuado para esta población mediante la valoración de necesidad de preparación del cuero cabelludo, rapidez de colocación, si se trataba de una aplicación invasiva, posibilidad de reposicionamiento, riesgo de lesiones cutáneas, esterilidad de la técnica y durabilidad. El electrodo elegido se utilizó para la monitorización continua electroencefalográfica iniciada en las primeras 24h de vida y mantenida al menos hasta las 72h.

Resultados:

Los electrodos evaluados fueron agujas subdérmicas, cucharillas de plata y 2 tipos de electrodos autoadhesivos (de hidrogel sólido y de gel conductor líquido). Los electrodos de gel conductor líquido fueron los elegidos. Se utilizaron en 41 neonatos con una edad gestacional media de 25,8±1,1 semanas. Se obtuvo una buena impedancia duradera de forma rápida y sin necesidad de manipulaciones excesivas y no observamos lesiones cutáneas. La satisfacción del personal involucrado en su colocación fue muy elevada. (AU)
ABSTRACT

Introduction:

Monitoring of brain function using continuous electroencephalography (aEEG/cEEG) is an essential tool in the standard care of the term infant, and its use is growing in the premature infant as a biomarker of lesion and brain maturity. However, the placing of the electrodes is a great challenge, particularly in the extremely premature infant, which often discourages neuromonitoring. The aim of this study is to assess the different electrodes available, to select the one that best suits the peculiarities of the extremely premature infant, and evaluate its applicability in clinical practice.Population and

methods:

With the aim of designing a neuromonitoring study protocol using aEEG/cEEG in<28 weeks premature infants, an analysis was made of our experience with the type of electrodes available. The electrode that was considered most suitable for this population was chosen by assessing the need of preparing the scalp, speed in positioning the electrodes, if the application was invasive or not, the possibility of repositioning, risk of skin injuries, sterility of the technique, and durability. The electrode chosen was used for continuous electroencephalographic monitoring started in the first 24h of life, and maintained until at least 72h of life.

Results:

The electrodes evaluated were subdermal needles, silver cups, and 2 types of self-adhesive electrodes (solid hydrogel and wet gel). The wet gel electrodes were chosen. They were used on 41 neonates with a mean gestational age of 25.8±1.1 weeks. Good stable impedance was rapidly obtained, without the need of excessive manipulations, and no skin injuries were observed. The satisfaction of the staff involved in positioning them was very high. (AU)
Assuntos


Texto completo: Disponível Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Impedância Elétrica / 34628 / Eletrodos / Lactente Extremamente Prematuro Limite: Humanos / Recém-Nascido Idioma: Espanhol Revista: An. pediatr. (2003. Ed. impr.) Ano de publicação: 2021 Tipo de documento: Artigo Instituição/País de afiliação: Hospital Universitario La Paz/España

Texto completo: Disponível Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Impedância Elétrica / 34628 / Eletrodos / Lactente Extremamente Prematuro Limite: Humanos / Recém-Nascido Idioma: Espanhol Revista: An. pediatr. (2003. Ed. impr.) Ano de publicação: 2021 Tipo de documento: Artigo Instituição/País de afiliação: Hospital Universitario La Paz/España
...