Cuantificación del efecto del movimiento vertical del paciente durante la gated-SPECT de perfusión miocárdica sobre los índices de perfusión y de función del ventrículo izquierdo / Quantifying the effect of vertical motion of patient during gated myocardial perfusion SPECT on indices of perfusion and function of left ventricle
Ahmadi, R; Qutbi, M.
Afiliação
  • Ahmadi, R; Universidad de Ciencias Médicas de Hamadán. Facultad de Medicina. Centro Cardiológico Farshchian. Hamadán. Irán
  • Qutbi, M; Universidad de Ciencias Médicas Shahid Beheshti. Facultad de Medicina. Hospital Universitario Taleghani. Teherán. Irán
Rev. esp. med. nucl. imagen mol. (Ed. impr.) ; 42(4): 238-248, jul.- ago. 2023. ilus, tab
Artigo em Espanhol | IBECS | ID: ibc-223280
Biblioteca responsável: ES1.1
Localização: ES15.1 - BNCS
RESUMEN
Objetivo Evaluar el efecto del movimiento de «rebote» simulado por ordenador en los índices de perfusión y función del ventrículo izquierdo en relación con 3determinantes principales del movimiento (duración, magnitud y tiempo) mediante la evaluación del efecto único y la interacción de estos atributos en un modelo estadístico. Métodos Para el estudio, se seleccionaron 29 exploraciones gated-SPECT de perfusión miocárdica y, a continuación, se simuló manualmente el patrón de movimiento de «rebote» con respecto a 3atributos principales del movimiento duración (corta vs. larga), magnitud (2 vs. 4 píxeles) y tiempo (precoz vs. tardío), todo ello en la dirección vertical ascendente. Todas las imágenes SPECT se reconstruyeron y filtraron utilizando un método idéntico (algoritmo OSEM) y con los mismos parámetros. Los índices de perfusión y función miocárdica del ventrículo izquierdo se obtuvieron a partir del programa informático QGS del Cedars-Sinai en las imágenes originales y de movimiento simulado y se comparan entre sí. Se aplicaron pruebas ANOVA de medidas repetidas intrasujeto de 2y 3vías para evaluar el efecto principal de cada variable o atributo y la interacción entre ellas. Resultados Las puntuaciones sumadas aumentaron aproximadamente de manera exponencial desde «sin movimiento» hasta un rebote corto y, después, hasta un rebote largo. En el rebote largo de 4 píxeles, los defectos de perfusión fueron notables. Todas las comparaciones de la extensión del defecto y del déficit total de perfusión fueron estadísticamente significativas. La diferencia media entre los patrones de movimiento de rebote corto y «sin movimiento» fue pequeña, incluso en los movimientos de 4 píxeles (casi del 3% o menor). Por el contrario, la diferencia media entre los patrones de movimiento de rebote largo y «sin movimiento» fue superior al 5% (AU)
ABSTRACT
Objective To assess the effect of software-simulated «bouncing» motion on left ventricle (LV) perfusion and function indices concerning 3main determinants of motion (duration, magnitude and time) by evaluating the sole effect and interaction of these attributes in a statistical model. Methods Twenty-nine gated myocardial perfusion SPECT scans were selected for the study and then, «bounce» motion pattern was simulated manually regarding 3main attributes of motion including duration (short versus long), magnitude (2 versus 4 pixels) and time (early versus late), all in upward vertical direction. All SPECT images are reconstructed and filtered with an identical method (OSEM algorithm) and same parameters. Indices of LV myocardial perfusion and function are derived using QGS package of Cedars-Sinai software in original and simulated-motion images and are then compared with each other. Two- and 3-way repeated measure within-subjects ANOVA tests are conducted to evaluate the main effect of each variable or attribute and the interaction between them. Results Summed scores increase roughly exponentially from «no motion» to short bounce and then, to long bounce. In long 4-pixel bounce, perfusion defects are remarkable. All comparisons of defect extent and total perfusion deficit are statistically significant. Mean difference between short bounce motion patterns with «no motion» is small even in 4-pixel movements (almost 3% or lower). In contrast, mean difference between long bounce motion patterns with «no motion» is higher than 5%. Using a paired-sample t test, in all pairs, mean difference for ejection fraction is less than 4% which all are statistically significant (AU)
Assuntos

Buscar no Google
Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Função Ventricular Esquerda / Imagem de Perfusão do Miocárdio / Movimento Limite: Humanos Idioma: Espanhol Revista: Rev. esp. med. nucl. imagen mol. (Ed. impr.) Ano de publicação: 2023 Tipo de documento: Artigo Instituição/País de afiliação: Universidad de Ciencias Médicas Shahid Beheshti/Irán / Universidad de Ciencias Médicas de Hamadán/Irán
Buscar no Google
Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Função Ventricular Esquerda / Imagem de Perfusão do Miocárdio / Movimento Limite: Humanos Idioma: Espanhol Revista: Rev. esp. med. nucl. imagen mol. (Ed. impr.) Ano de publicação: 2023 Tipo de documento: Artigo Instituição/País de afiliação: Universidad de Ciencias Médicas Shahid Beheshti/Irán / Universidad de Ciencias Médicas de Hamadán/Irán
...