Levantamento das condições de funcionamento dos serviços de radiologia de hospitais públicos e universitários do Rio de Janeiro / Evaluation of working conditions of radiology departments of public and university hospitals in Rio de Janeiro, Brazil
Radiol. bras
; 37(4): 271-278, jul.-ago. 2004. ilus, tab, graf
Artigo
em Português
| LILACS
| ID: lil-364712
Biblioteca responsável:
BR734.1
RESUMO
Foram realizados testes de controle de qualidade, descritos na Portaria 453/98 do Ministério da Saúde, em sete hospitais públicos do Rio de Janeiro, em um total de 29 tubos de raios X, 30 chassis e 20 vestimentas de proteção individual. Os testes avaliaram o desempenho do gerador de raios X e a geometria do feixe. Foram verificados, também, chassis, "écrans" e negatoscópios, assim como as câmaras claras e escuras. Alguns dos resultados encontrados foram 55 por cento dos tubos foram reprovados em exatidão e 15 por cento em reprodutibilidade de kVp; 53 por cento foram reprovados em exatidão e 31 por cento em reprodutibilidade de tempo de exposição; 30 por cento foram reprovados em camada semi-redutora; 18 por cento foram reprovados no alinhamento do feixe central; 30 por cento foram reprovados em coincidência de campos; 40 por cento foram reprovados na linearidade de exposição; 64 por cento foram rejeitados no teste de ponto focal. Foram verificadas a guarda inadequada e a falta de recomendação para uso dos aventais plumbíferos. Os resultados dos testes sugerem a necessidade da implementação sistemática de controle de qualidade por um profissional qualificado. A relação de custo-benefício reforça a implantação e a manutenção dos procedimentos contidos e executados neste trabalho, pois tais procedimentos garantiriam a diminuição do custo final do serviço e da dose nos pacientes.
ABSTRACT
Quality control tests, according to "Portaria 453/98" of the Brazilian Health Ministry, were applied in seven public hospitals of Rio de Janeiro including 29 X-ray tubes, 30 cassettes and 20 individual protection garments. These tests evaluate mainly the X-ray generator and system geometry. Other tests in the accessories were also performed, such as cassettes, screens, viewing boxes, darkrooms and clear rooms. In the kVp tests, 55% of the equipments were disapproved in exactitude and 15% in reproducibility. Time exactitude and reproducibility were not approved in 53% and 31% of the equipments, respectively. The percentages of disapprovement were for the HVL (half-value layer) 30%, for beam alignment 18%, for field coincidence 30%, for exposure linearity 40% and for focal spot 64%. It has also been observed that frequently the lead aprons were not kept in appropriate hangers and this procedure may cause damage to these accessories. The results of the test suggest the need of implementation of quality assurance tests in these services. The cost benefit analysis of the results reinforces the importance of the implementation of the recommendations contained in this paper.
Texto completo:
Disponível
Coleções:
Bases de dados internacionais
Contexto em Saúde:
Agenda de Saúde Sustentável para as Américas
Problema de saúde:
Objetivo 6: Sistemas de informação em saúde
Base de dados:
LILACS
Assunto principal:
Serviço Hospitalar de Radiologia
/
Sistemas de Informação em Radiologia
Tipo de estudo:
Estudo diagnóstico
/
Estudo de avaliação
/
Guia de prática clínica
Limite:
Humanos
País/Região como assunto:
América do Sul
/
Brasil
Idioma:
Português
Revista:
Radiol. bras
Assunto da revista:
Diagnóstico por Imagem
/
Radiologia
Ano de publicação:
2004
Tipo de documento:
Artigo
País de afiliação:
Brasil
Instituição/País de afiliação:
Fundação Oswaldo Cruz/BR
/
Universidade Federal do Rio de Janeiro/BR