Your browser doesn't support javascript.
loading
Seguimiento en fase aguda de catpeteres electrodos de estimulación endocavitaria permanente: Comparación de tres modelos / Comparison of three brands of intracardiac pacemarker leads
Frangini S., Patricia; Vergara S., Ismael; González A., Rolando; Fajuri N., Alejandro; Baeza L., Mariana.
Afiliação
  • Frangini S., Patricia; Pontificia Universidad Católica de Chile. Facultad de Medicina. Departamento de Enfermedades Cardiovasculares. Laboratorio de Electrofisiología Cardíaca. Santiago. CL
  • Vergara S., Ismael; Pontificia Universidad Católica de Chile. Facultad de Medicina. Departamento de Enfermedades Cardiovasculares. Laboratorio de Electrofisiología Cardíaca. Santiago. CL
  • González A., Rolando; Pontificia Universidad Católica de Chile. Facultad de Medicina. Departamento de Enfermedades Cardiovasculares. Laboratorio de Electrofisiología Cardíaca. Santiago. CL
  • Fajuri N., Alejandro; Pontificia Universidad Católica de Chile. Facultad de Medicina. Departamento de Enfermedades Cardiovasculares. Laboratorio de Electrofisiología Cardíaca. Santiago. CL
  • Baeza L., Mariana; Pontificia Universidad Católica de Chile. Facultad de Medicina. Departamento de Enfermedades Cardiovasculares. Laboratorio de Electrofisiología Cardíaca. Santiago. CL
Rev. méd. Chile ; 134(11): 1427-1435, nov. 2006. ilus, tab
Article em Es | LILACS | ID: lil-439938
Biblioteca responsável: BR1.1
ABSTRACT

Background:

Electrode lead design and materials influence their performance, stability and manipulation characteristics. In our laboratory, we use straight intracardiac, active fixation, steroid eluting leads. These features are shared by three brands of pacemarker distributors.

Aim:

To compare the short term results of three brands of leads used in our laboratory in patients requiring the implant of a pacemarker of cardioverter. Material and

methods:

One hundred and four patients (mean age 70 years, 59 males) subjected to a pacemarker or cardioverter implant were studied and followed during the first three months post implant. In these patients, 49 Guidant Flextend® 4087 or 4088, 27 Saint Jude Tendril® 1488T and 10 Medtronic Capsurefix® 5076 leads were implanted in the right atrium and 60 Guidant Flextend® 4087 or 4088, 29 Saint Jude Tendril® 1488T and 19 Medtronic Capsurefix® 5076 leads were implanted in the right ventricle.

Results:

Implant parameters were adequate for all leads. A sub-acute rise in ventricular stimulation threshold was detected in one Flextrend® lead. Three atrial leads (two Flextend® and one Capsurefix®) and one Capsurefix® ventricular lead experienced an acute displacement. One patient with a Flextend® lead, had a cardiac tamponade caused by an atrial perforation.

Conclusions:

The three brands of leads tested can be successfully implanted with comparable parameters and without differences in the evolution of patients during the first three months.
Assuntos
Texto completo: 1 Coleções: 01-internacional Base de dados: LILACS Assunto principal: Marca-Passo Artificial / Arritmias Cardíacas / Desfibriladores Implantáveis Tipo de estudo: Observational_studies / Prognostic_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male Idioma: Es Revista: Rev. méd. Chile Assunto da revista: MEDICINA Ano de publicação: 2006 Tipo de documento: Article País de afiliação: Chile
Texto completo: 1 Coleções: 01-internacional Base de dados: LILACS Assunto principal: Marca-Passo Artificial / Arritmias Cardíacas / Desfibriladores Implantáveis Tipo de estudo: Observational_studies / Prognostic_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male Idioma: Es Revista: Rev. méd. Chile Assunto da revista: MEDICINA Ano de publicação: 2006 Tipo de documento: Article País de afiliação: Chile
...