Your browser doesn't support javascript.
loading
Extensometria: estudo das microdeformações ao redor de três implantes de conexão protética cone-morse, fundidos com coifas plásticas e usinadas, submetidos à carga vertical estática / A comparative study of microstrain around three-morse taper implants,casting with machined and plastic copings, under axial loading
Nishioka, Renato Sussumu; Santos, Vivian Mayumi Miyazaki; Nishioka, Gabriela Nogueira de Melo; Balducci, Ivan.
Afiliação
  • Nishioka, Renato Sussumu; Universidade Estadual Paulista. Faculdade de Odontologia. Departamento de Materiais Odontológicos e Prótese. São José dos Campos. BR
  • Santos, Vivian Mayumi Miyazaki; Universidade Estadual Paulista. Faculdade de Odontologia. Programa de Pós-graduação em Odontologia Restauradora. São José dos Campos. BR
  • Nishioka, Gabriela Nogueira de Melo; Universidade Estadual Paulista. Faculdade de Odontologia. São José dos Campos. BR
  • Balducci, Ivan; Universidade Estadual Paulista. Faculdade de Odontologia. Departamento de Odontologia Social e Clínica Infantil. São José dos Campos. BR
Rev. odontol. UNESP (Online) ; 41(1): 54-59, jan.-fev. 2012. ilus, tab
Artigo em Português | LILACS, BBO - Odontologia | ID: lil-667006
Biblioteca responsável: BR39.2
RESUMO

Objetivo:

Analisar, por meio da extensometria, as microdeformações ao redor de implantes cone-Morse com diferentes tipos de coifas-plásticas (n = 5) e usinadas (n = 5). Foram inseridos linearmente três implantes cone-Morse no bloco de poliuretano.

Método:

Quatro extensômetros (SG) foram colados na superfície do poliuretano e foram identificados pelo SG1, que foi colado adjacente e mesialmente ao implante A, pelos SG2 e SG3 colados adjacentes mesial e distalmente ao implante B, e o SG4, que foi colado distalmente ao implante C. Dez supraestruturas em CoCr foram parafusadas sobre pilares protéticos Microunit com torque de 10 N.cm, com o auxílio de um torquímetro manual. Cargas axiais de 30 kg foram aplicadas em pontos pré-determinados (A, B, C) sobre os respectivos parafusos de retenção das próteses. Os extensômetros foram capazes de analisar as microdeformações em um limitado segmento ao redor dos implantes. Os registros foram comparados utilizando-se ANOVA de medidas repetitivas e teste múltiplo de Tukey (a = 0,005).

Resultado:

Os registros não evidenciaram diferença estatística signficante para o tipo de coifa (p = 0,0838); contudo, evidenciou-se diferença estatística para os pontos de aplicação de carga (p = 0,0338).

Conclusão:

Não houve evidências de que as coifas usinadas reduziram as microdeformações ao redor dos implantes. O ponto de aplicação de carga determinou a quantidade de microdeformações.
ABSTRACT

Objective:

The aim of this in vitro study was quantify the microstrain around Morse taper implants using plastic (n = 5) and machined (n = 5) copings under axial loading.

Methods:

Three implants of Morse taper junction were embedded in the center of the polyurethane block. Four Strain gauges (SG) were bonded on the surface of polyurethane and it was designated SG1 placed mesially adjacent to implant A, SG2 and SG3 were placed mesially and distally adjacent to the implant B and SG4 was placed distally adjacent to the implant C. The ten CoCr superstructure?s occlusal screws were tightened onto the Microunit abutments with a torque of 10 N.cm using the manufacture?s manual torque- controlling device. Axial load of 30 kg was applied at three predetermined points (A, B, C). The strain gauges were only capable of detecting strain in a limited segment around the implants. All of the data were compared using RM ANOVA and post hoc multiple range Tukey test (a = 0.005). Result. The results did not show a statistically significant difference for the copings (p = 0.0838), but showed a statistically significant difference for the loading points (p = 0.0338).

Conclusion:

There were not evidences that the machined copings in reducing the strain around implant. The loading point persuaded in the magnitude of micro strain.
Assuntos

Texto completo: Disponível Coleções: Bases de dados internacionais Base de dados: BBO - Odontologia / LILACS Assunto principal: Próteses e Implantes / Fenômenos Biomecânicos / Análise de Variância / Suporte de Carga / Implantação Dentária Endóssea / Projeto do Implante Dentário-Pivô Idioma: Português Revista: Rev. odontol. UNESP (Online) Assunto da revista: Odontologia Ano de publicação: 2012 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: Universidade Estadual Paulista/BR
Texto completo: Disponível Coleções: Bases de dados internacionais Base de dados: BBO - Odontologia / LILACS Assunto principal: Próteses e Implantes / Fenômenos Biomecânicos / Análise de Variância / Suporte de Carga / Implantação Dentária Endóssea / Projeto do Implante Dentário-Pivô Idioma: Português Revista: Rev. odontol. UNESP (Online) Assunto da revista: Odontologia Ano de publicação: 2012 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: Universidade Estadual Paulista/BR
...