Condutividade e teste quantitativo coulométrico na triagem neonatal para fibrose cística / Sweat conductivity and coulometric quantitative test in neonatal cystic fibrosis screening
Domingos, Mouseline Torquato; Magdalena, Neiva Isabel Rodrigues; Cat, Mônica Nunes Lima; Watanabe, Alexandra Mitiru; Rosário Filho, Nelson Augusto.
Afiliação
  • Domingos, Mouseline Torquato; Fundação Ecumênica de Proteção ao Excepcional. Serviço de Referência em Triagem Neonatal. Curitiba. BR
  • Magdalena, Neiva Isabel Rodrigues; Fundação Ecumênica de Proteção ao Excepcional. Serviço de Referência em Triagem Neonatal. Curitiba. BR
  • Cat, Mônica Nunes Lima; Fundação Ecumênica de Proteção ao Excepcional. Serviço de Referência em Triagem Neonatal. Curitiba. BR
  • Watanabe, Alexandra Mitiru; Fundação Ecumênica de Proteção ao Excepcional. Serviço de Referência em Triagem Neonatal. Curitiba. BR
  • Rosário Filho, Nelson Augusto; Fundação Ecumênica de Proteção ao Excepcional. Serviço de Referência em Triagem Neonatal. Curitiba. BR
J. pediatr. (Rio J.) ; 91(6): 590-595, nov.-dez. 2015. tab, graf
Artigo em Inglês | LILACS | ID: lil-769800
Biblioteca responsável: BR1.1
RESUMO
Resumo Objetivo Comparar os resultados obtidos no teste do suor pelo método da condutividade e a dosagem coulométrica de cloreto no suor em recém0nascidos (RN) suspeitos da triagem neonatal para fibrose cística (FC). Métodos O teste do suor foi feito simultaneamente pelos dois métodos em crianças com e sem FC. Os valores de corte para confirmar FC foram na condutividade > 50 mmol/L e no teste coulométrico > 60 mmol/L. Resultados Fizeram o teste do suor por condutividade e dosagem coulométrica simultaneamente 444 RN sem FC (185 do sexo masculino, 234 do feminino e 24 não informado) e obtiveram resultado mediano de 32 mmol/L e 12 mmol/L respectivamente. Para os noventa RN com FC os valores medianos de condutividade e dosagem coulométrica foram 108 mmol/L e 97 mmol/L respectivamente. O índice de falso positivo para condutividade foi de 16,7% e em todos os pacientes FC foi superior a 50 mmol/L, o que confere ao método 100% de sensibilidade (IC 95% = 93,8 a 97,8), especificidade de 96,2% (IC 95% = 93,8 a 97,8), valor preditivo positivo 83,3 (IC 95% = 74,4 a 91,1), valor preditivo negativo 100% (IC 95% = 90,5 a 109,4) e acurácia 9,8%. A correlação entre os métodos foi de r = 0,97 (p > 0,001).O melhor valor de corte sugerido foi de 69,0 mmol/L, coeficiente de kappa = 0,89. Conclusão O teste da condutividade apresentou excelente correlação com o quantitativo coulométrico, alta sensibilidade e especificidade e pode ser usado no diagnóstico da FC em crianças detectadas pela triagem neonatal.
ABSTRACT
Abstract Objective To compare the results obtained with the sweat test using the conductivity method and coulometric measurement of sweat chloride in newborns (NBs) with suspected cystic fibrosis (CF) in the neonatal screening program. Methods The sweat test was performed simultaneously by both methods in children with and without CF. The cutoff values to confirm CF were >50 mmol/L in the conductivity and >60 mmol/L in the coulometric test. Results There were 444 infants without CF (185 males, 234 females, and 24 unreported) submitted to the sweat test through conductivity and coulometric measurement simultaneously, obtaining median results of 32 mmol/L and 12 mmol/L, respectively. For 90 infants with CF, the median values of conductivity and coulometric measurement were 108 mmol/L and 97 mmol/L, respectively. The false positive rate for conductivity was 16.7%, and was higher than 50 mmol/L in all patients with CF, which gives this method a sensitivity of 100% (95% CI 93.8-97.8), specificity of 96.2% (95% CI 93.8-97.8), positive predictive value of 83.3% (95% CI 74.4-91.1), negative predictive value of 100% (95% CI 90.5-109.4), and 9.8% accuracy. The correlation between the methods was r = 0.97 (p > 0.001). The best suggested cutoff value was 69.0 mmol/L, with a kappa coefficient = 0.89. Conclusion The conductivity test showed excellent correlation with the quantitative coulometric test, high sensitivity and specificity, and can be used in the diagnosis of CF in children detected through newborn screening.
Assuntos


Texto completo: Disponível Coleções: Bases de dados internacionais Base de dados: LILACS Assunto principal: Suor / Cloretos / Fibrose Cística Tipo de estudo: Estudo diagnóstico / Estudo observacional / Estudo de prevalência / Estudo prognóstico / Fatores de risco / Estudo de rastreamento Limite: Feminino / Humanos / Masculino / Recém-Nascido Idioma: Inglês Revista: J. pediatr. (Rio J.) Assunto da revista: Pediatria Ano de publicação: 2015 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: Fundação Ecumênica de Proteção ao Excepcional/BR

Texto completo: Disponível Coleções: Bases de dados internacionais Base de dados: LILACS Assunto principal: Suor / Cloretos / Fibrose Cística Tipo de estudo: Estudo diagnóstico / Estudo observacional / Estudo de prevalência / Estudo prognóstico / Fatores de risco / Estudo de rastreamento Limite: Feminino / Humanos / Masculino / Recém-Nascido Idioma: Inglês Revista: J. pediatr. (Rio J.) Assunto da revista: Pediatria Ano de publicação: 2015 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: Fundação Ecumênica de Proteção ao Excepcional/BR
...