Evaluation of multielement catheter-cooled interstitial ultrasound applicators for high-temperature thermal therapy.
Nau, W H; Diederich, C J; Burdette, E C.
Afiliação
  • Nau WH; Department of Radiation Oncology, University of California, San Francisco 94143-0226, USA. nau@itsa.ucsf.edu
Med Phys ; 28(7): 1525-34, 2001 Jul.
Article em En | MEDLINE | ID: mdl-11488586
ABSTRACT
Catheter-cooled (CC) interstitial ultrasound applicators were evaluated for their use in high-temperature coagulative thermal therapy of tissue. Studies in ex vivo beef muscle were conducted to determine the influences of applied electrical power levels (5-20 W per element), catheter flow rate (20-60 ml min(-1)), circulating water temperature (7-40 degrees C), and frequency (7-9 MHz) on temperature distribution and thermal lesion geometry. The feasibility of using multiple interstitial applicators to thermally coagulate a predetermined volume of tissue was also investigated. Results of these studies revealed that the directional shape of the thermal lesions is maintained with increasing time and power. Radial depths of the thermal lesions ranged from 10.7 +/- 0.7 mm after heating for 4 min with an applied power level of 5 W, to 16.2 +/- 1.4 mm with 20 W. The axial length of the thermal lesions is controlled tightly by the number of active transducers. A catheter flow rate of 20 to 40 ml min(-1) (52.2 +/- 5.5 kPa at 40 ml min(-1)) with 22 degrees C water was determined to provide sufficient cooling of the transducers for power levels used in this study. In vivo temperatures measured in the center of a 3-cm-diam peripheral implant of four applicators in pig thigh muscle reached 89.3 degrees C after 4 min of heating, with boundaries of coagulation clearly defined by applicator position and directivity. Conformability of heating in a clinically relevant model was demonstrated by inserting two directional CC applicators with a 2 cm separation within an in vivo canine prostate, and generating a thermal lesion measuring 3.8 cm x 2.2 cm in cross section while directing energy away from, and protecting the rectum. Maximum measured temperatures at midgland exceeded 90 degrees C within 20 min of heating. The results of this study demonstrate the utility of single or multiple CC applicators for conformal thermal coagulation and high temperature thermal therapy, with potential for clinical applications in sites such as prostate, liver, breast, or uterus.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Transdutores / Terapia por Ultrassom / Cateterismo / Temperatura Alta Tipo de estudo: Guideline Limite: Animals Idioma: En Revista: Med Phys Ano de publicação: 2001 Tipo de documento: Article País de afiliação: Estados Unidos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Transdutores / Terapia por Ultrassom / Cateterismo / Temperatura Alta Tipo de estudo: Guideline Limite: Animals Idioma: En Revista: Med Phys Ano de publicação: 2001 Tipo de documento: Article País de afiliação: Estados Unidos