Your browser doesn't support javascript.
loading
Electrochemiluminescent determination of L-cysteine with a flow-injection analysis system.
Zhu, Liande; Li, Yingxiu; Zhu, Guoyi.
Afiliação
  • Zhu L; State Key Laboratory of Electroanalytical Chemistry, Changchun Institute of Applied Chemistry, Chinese Academy of Sciences, Changchun 130022, PR China.
Anal Sci ; 19(4): 575-8, 2003 Apr.
Article em En | MEDLINE | ID: mdl-12725394
ABSTRACT
A flow-injection electrochemiluminescent method for L-cysteine determination has been developed based on its enhancement of the electrochemiluminecence of luminol at a glassy carbon electrode. This method is simple and sensitive for cysteine determination. Under the selected experimental parameters, the linear range for cysteine concentration was 1.0 x 10(-6) - 5.0 x 10(-5) mol/l, and the detection limit was 0.67 micromol/l (S/N = 3). The relative standard deviation for 11 measurements of 1.0 x 10(-5) mol/l cysteine was 4.5%. The proposed method has been applied to the detection of cysteine in pharmaceutical injections with satisfactory results.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Análise de Injeção de Fluxo / Cisteína / Eletroquímica Tipo de estudo: Diagnostic_studies Idioma: En Revista: Anal Sci Ano de publicação: 2003 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Análise de Injeção de Fluxo / Cisteína / Eletroquímica Tipo de estudo: Diagnostic_studies Idioma: En Revista: Anal Sci Ano de publicação: 2003 Tipo de documento: Article