This paper describes a stack phantom useful for imaging complex activity distributions. It is based on images printed with radioactive ink using a commercial ink-jet printer. The application for the phantom is in the evaluation of planar and SPECT scintillation camera images and for validation of Monte Carlo simulated images. The accuracy in generating the activity distributions on paper sheets is especially important. Here we describe the calibration procedure for the ink-jet printer. The goal of the printer calibration is to find the relationship between the digital image count (voxel grey level) and its corresponding activity on the paper sheets (radioactivity). The relationship between the voxel grey level and the radioactivity on the paper sheets (measured by scanning technique and well counter) was found to be logarithmic, and a 3rd degree polynomial was found to fit the relationship. The distribution of radioactivity in the ink cartridge was investigated by pinhole SPECT. The distribution of (99m)Tc solution was found to be homogeneous in the ink solution. Experimental studies were done directly on Monte Carlo simulated heart images from the NCAT phantom. The result showed that the simulated images are similar to the images measured using the ink-jet technique. This stack phantom could be a promising solution with an advantage that the exact geometry generated in Monte Carlo could be imitated in the phantom. The phantom is a very flexible device and clearly much more versatile than conventional phantoms which have a fixed geometry and spatial limitation.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tomografia Computadorizada de Emissão de Fóton Único / Imagens de Fantasmas / Miocárdio / Medicina Nuclear Tipo de estudo: Diagnostic_studies / Health_economic_evaluation / Prognostic_studies Limite: Humans Idioma: En Revista: Cancer Biother Radiopharm Assunto da revista: FARMACIA / FARMACOLOGIA / NEOPLASIAS / TERAPEUTICA Ano de publicação: 2003 Tipo de documento: Article País de afiliação: Suécia País de publicação: Estados Unidos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tomografia Computadorizada de Emissão de Fóton Único / Imagens de Fantasmas / Miocárdio / Medicina Nuclear Tipo de estudo: Diagnostic_studies / Health_economic_evaluation / Prognostic_studies Limite: Humans Idioma: En Revista: Cancer Biother Radiopharm Assunto da revista: FARMACIA / FARMACOLOGIA / NEOPLASIAS / TERAPEUTICA Ano de publicação: 2003 Tipo de documento: Article País de afiliação: Suécia País de publicação: Estados Unidos