Stereotactic radiosurgery planning with ictal SPECT images.
Australas Phys Eng Sci Med
; 27(3): 136-47, 2004 Sep.
Article
em En
| MEDLINE
| ID: mdl-15580843
This paper is motivated by a clinical requirement to utilise ictal SPECT images for target localisation in stereotactic radiosurgery treatment planning using the xknife system which only supports CT and MRI images. To achieve this, the SPECT images were converted from raw (pixel data only) format into a part 10 compliant DICOM CT fileset. The minimum requirements for the recasting of a raw format image as DICOM CT or MRI data set are described in detail. The method can be applied to the importation of raw format images into any radiotherapy treatment planning system that supports CT or MRI import. It is demonstrated that the combination of the low spatial resolution SPECT images, depicting functional information, with high spatial resolution MRI images, which show the structural information, is suitable for stereotactic radiosurgery treatment planning.
Buscar no Google
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Radioterapia Assistida por Computador
/
Interpretação de Imagem Assistida por Computador
/
Tomografia Computadorizada de Emissão de Fóton Único
/
Radiocirurgia
/
Cirurgia Assistida por Computador
/
Epilepsia
Tipo de estudo:
Diagnostic_studies
/
Evaluation_studies
Limite:
Humans
Idioma:
En
Revista:
Australas Phys Eng Sci Med
Ano de publicação:
2004
Tipo de documento:
Article
País de afiliação:
Austrália
País de publicação:
Holanda