Novel approach for assessing performance of PCR cyclers used for diagnostic testing.
Schoder, D; Schmalwieser, A; Schauberger, G; Hoorfar, J; Kuhn, M; Wagner, M.
Afiliação
  • Schoder D; Institute for Milk Hygiene, Milk Technology and Food Science, Veterinary University, Vienna, Austria.
J Clin Microbiol ; 43(6): 2724-8, 2005 Jun.
Article em En | MEDLINE | ID: mdl-15956389
As part of a large international project for validation and standardization of PCR, the influence of thermocyclers on PCR was tested. Six brand-new, Peltier technology-driven 96-well thermocyclers were subjected to a novel and stringent in-tube (not block) physical testing. The temperature was directly monitored in PCR tubes containing 50 microl of distilled water at 13 different block positions. The certified temperature accuracy of the measurement system was +/-0.3 degrees C. Finally, the results of the physical testing were compared to those of an amplification efficiency study running an in-house PCR assay. The cyclers did not perform within the manufacturer's specification. Premature timing, under- and overshooting, and spatial variation of heat transfer were found to be the critical factors. The physical testing allowed us to distinguish accurate from less-accurate (2/6) cyclers. The lack of thermal homogeneities became most evident at the denaturation level during the first 15 s. At the time point zero, the accurate cyclers showed temperature deviations of 0.5 to 1.5 degrees C, whereas less-accurate cyclers failed to reach the set temperature by 13 to 20 degrees C. Consequently, the two less-accurate cyclers could not gain positive PCR results by running an in-house PCR assay. However, by modifying the original temperature protocol by increasing the denaturation temperature and time, the amplification efficiency of these two cyclers could be improved significantly. The results have implication for laboratories using diagnostic PCR testing.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Reação em Cadeia da Polimerase Tipo de estudo: Diagnostic_studies / Evaluation_studies / Guideline Idioma: En Revista: J Clin Microbiol Ano de publicação: 2005 Tipo de documento: Article País de afiliação: Áustria País de publicação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Reação em Cadeia da Polimerase Tipo de estudo: Diagnostic_studies / Evaluation_studies / Guideline Idioma: En Revista: J Clin Microbiol Ano de publicação: 2005 Tipo de documento: Article País de afiliação: Áustria País de publicação: Estados Unidos