A hybrid approach for quantification of aortic valve stenosis using cardiac magnetic resonance imaging and echocardiography: comparison to right heart catheterization and standard echocardiography.
Clin Res Cardiol
; 95(3): 162-7, 2006 Mar.
Article
em En
| MEDLINE
| ID: mdl-16598529
We replaced Doppler-derived stroke volume in the continuity equation (method A) by either right heart catheterization-derived stroke volume (method B) or cardiovascular magnetic resonance-derived stroke volume (method C) to calculate aortic valve area in 20 consecutive patients with moderate or severe aortic stenosis. Comparison of both hybrid methods (methods B and C) by Bland-Altman analysis showed a mean difference near zero, a spread within two standard deviations and very similar limits of agreement. More importantly, all patients were classified into the same category of severity by both methods.