Intercomparison of dosimetry systems based on CaF2:Mn TL detectors.
Radiat Prot Dosimetry
; 119(1-4): 300-5, 2006.
Article
em En
| MEDLINE
| ID: mdl-16735564
ABSTRACT
The responses of readings by the TL dosimetry system MR200 TL developed in-house and used at JSI and the TOLEDO TL system used at RBI are compared. Ten measurements at different doses ranging from 0.01 mSv to 5 Sv were carried out. A set of 36 dosemeters with three pellets of CaF2Mn were irradiated in radiation fields of 137Cs and 60Co. Analysis of the measured results shows that at doses below 0.1 Sv, readers' outputs do not differ >5% from each other. At doses >1 Sv, the results obtained by the MR200 reader must be corrected with a known factor. Finally, the reproducibility of the results from the MR200 was tested.