ABSTRACT
OBJECTIVES:
To investigate whether patency of a thin walled 8 mm fluoropassivated Dacron graft was similar to that of a standard 8mm PTFE graft for femorofemoral crossover bypass surgery.DESIGN:
A randomised multicentre clinical trial comparing two vascular grafts with participation of 10 departments of vascular surgery in Denmark, Sweden and Norway. PATIENTS ANDMETHODS:
198 patients were randomised to PTFE (n=107) or fluoropolymer-coated Dacron grafts (n=91), 63% underwent surgery for claudication, 27% for ischaemic rest pain and 10% for tissue loss. The median follow-up time was 24 months (IQR 19-26 months).RESULTS:
The primary patency rate of the two grafts was similar (log rank test p=0.35). The primary patency rates (95% CI) for coated Dacron and PTFE grafts were 92% (86-98) and 94% (89-99) at 12 months and 87% (74-95) and 93% (87-99) at 24 months, respectively.CONCLUSION:
In patients with unilateral iliac artery disease not amenable to angioplasty, the femoral-femoral bypass is durable and effective. No difference in patency was found between the two graft materials (fluoropolymer coated Dacron and PTFE).
Buscar no Google
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Politetrafluoretileno
/
Polímeros de Fluorcarboneto
/
Prótese Vascular
/
Polietilenotereftalatos
/
Implante de Prótese Vascular
/
Materiais Revestidos Biocompatíveis
/
Artéria Femoral
Tipo de estudo:
Clinical_trials
/
Etiology_studies
/
Risk_factors_studies
Limite:
Aged
/
Female
/
Humans
/
Male
Idioma:
En
Revista:
Eur J Vasc Endovasc Surg
Assunto da revista:
ANGIOLOGIA
Ano de publicação:
2006
Tipo de documento:
Article
País de afiliação:
Dinamarca