Effect of HbS in the determination of HbA2 with the Biorad Variant II analyzer.
Clin Biochem
; 40(9-10): 744-6, 2007 Jun.
Article
em En
| MEDLINE
| ID: mdl-17442293
ABSTRACT
OBJECTIVES:
The effect of the presence of HbS in the determination of HbA2 using the Biorad Variant II analyzer. DESIGN ANDMETHODS:
The effect of HbS presence in the samples was quantified using the HELENA SAS-MX alkaline gel electrophoresis kit as the reference method.RESULTS:
The %HbA2 values from the Variant II analyzer and the HELENA SAS-MX alkaline gel electrophoresis kit show a good linear correlation in the absence of HbS. A strong positive bias in the %HbA2 values from the Variant II is apparent in the presence of HbS in the samples, when compared to the alkaline electrophoresis gel.CONCLUSION:
The Variant II analyzer gives reliable results for %HbA2 determination when no HbS is detectable in the samples. When HbS is present, the gel electrophoresis method gives more accurate results.