Will the hold of solid biodegradable implants be influenced by swelling during the degradation process? An in-vitro study with Meniscus Arrows.
Wouters, Diederick B; Bos, Rudolf R M; De Hosson, Jeff T.
Afiliação
  • Wouters DB; Department of General and Arthroscopic Surgery and Traumatology, TweeSteden Hospital, Dr. Deelenlaan 5, 5042 AD, Tilburg, and Department of Oral and Maxillofacial Surgery, University Medical Centre, Groningen, The Netherlands. dwouters@tsz.nl
Knee Surg Sports Traumatol Arthrosc ; 15(10): 1204-9, 2007 Oct.
Article em En | MEDLINE | ID: mdl-17589827
ABSTRACT
Water uptake after implantation of biodegradable devices induces swelling, as mentioned in literature. The hold in bone of solid devices will increase if the swelling is substantial enough. The results of weighing six Meniscus Arrows (MAs) before and after immersion in a sterile phosphate buffered saline solution during different time intervals were compared with the outcome of measurements under a field emission scanning electron microscope of six other MAs, stored under comparable conditions. The data were statistically evaluated with the Wilcoxon's signed rank test. The weight increase of 2.1 mg or 9.16% was statistically significant in the first 2 h following immersion, remaining stable afterwards with an average weight gain of 1.7 mg or 7.18%. The core diameter of the MAs increased to 0.01 mm or 1.01% with time. Although this is statistically significant, it is not expected to have any consequences for the hold. However, a remarkable and statistically significant decrease in the outer inter-barb diameter of 0.15 mm or 8.6% was noted with time. Mechanical testing should reveal the clinical relevance of the results of this study.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Dispositivos de Fixação Ortopédica / Teste de Materiais / Implantes Absorvíveis / Imersão Limite: Humans Idioma: En Revista: Knee Surg Sports Traumatol Arthrosc Assunto da revista: MEDICINA ESPORTIVA / TRAUMATOLOGIA Ano de publicação: 2007 Tipo de documento: Article País de afiliação: Holanda
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Dispositivos de Fixação Ortopédica / Teste de Materiais / Implantes Absorvíveis / Imersão Limite: Humans Idioma: En Revista: Knee Surg Sports Traumatol Arthrosc Assunto da revista: MEDICINA ESPORTIVA / TRAUMATOLOGIA Ano de publicação: 2007 Tipo de documento: Article País de afiliação: Holanda