Determination of ammonium in a buddingtonite sample by ion-chromatography.
Klock, P R; Lamothe, P J.
Afiliação
  • Klock PR; U.S. Geological Survey, Menlo Park, CA 94025, USA.
Talanta ; 33(6): 495-8, 1986 Jun.
Article em En | MEDLINE | ID: mdl-18964130
ABSTRACT
An ion-chromatographic method for the direct determination of ammonium, potassium, and sodium in geologic materials is described. Samples are decomposed with a mixture of hydrofluoric and hydrochloric acids in a sealed polycarbonate bottle heated in a microwave oven. The ion-chromatograph separates the cations and determines them by conductivity measurement. The ammonium concentrations thus determined have been verified by use of an ammonia-specific electrode. A total of 32 analyses of ammonium salts by both techniques showed an average error of -4%, with a relative standard deviation (RSD) of 6%. The ammonium concentrations found in a buddingtonite sample had an RSD of 2.2% and their mean agreed with that obtained by the Kjeldahl method. By use of the prescribed dilution of the sample, detection limits of 0.1% can be achieved for all three cations.
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Talanta Ano de publicação: 1986 Tipo de documento: Article País de afiliação: Estados Unidos