Your browser doesn't support javascript.
loading
Wall shear stress in intracranial self-expanding stents studied using ultra-high-resolution 3D reconstructions.
Benndorf, G; Ionescu, M; Y Alvarado, M V; Hipp, J; Metcalfe, R.
Afiliação
  • Benndorf G; Department of Radiology, Baylor College of Medicine, Houston, TX 77030, USA. benndorf@bcm.tmc.edu
AJNR Am J Neuroradiol ; 30(3): 479-86, 2009 Mar.
Article em En | MEDLINE | ID: mdl-19039050
ABSTRACT
BACKGROUND AND

PURPOSE:

Imaging of intracranial stents is constrained by resolution limits of current clinical imaging techniques providing insufficient visualization of deployment details and impeding its use for computational hemodynamic (CHD) simulations. The purpose of our study was to evaluate whether ultra-high-resolution MicroCT scans can illuminate detailed aspects of realistic in vitro stent deployment and serve as a reliable basis for CHD simulations of blood flow through self-expanding intracranial stents. MATERIALS AND

METHODS:

A Neuroform Treo (NF) stent and an Enterprise (ENT) stent were deployed in identical straight polytetrafluoroethylene tubes filled with contrast agent. MicroCT scans were obtained at a spatial resolution of 14 mum and used for ultra-high-resolution 3D reconstructions. CHD simulations were performed, with particular emphasis on local flow behavior near the wall and struts. Flow differences between the geometrically different stents were studied.

RESULTS:

MicroCT data revealed strut prolapse near the markers for the closed-cell design (ENT) stent and at some of the unconnected vertices of the open-cell design (NF) stent, which also showed some misalignments. CHD simulations showed that reverse wall shear stress occurred near some of the strut vertices and markers for the NF but only near the markers for the ENT.

CONCLUSIONS:

This work demonstrates the feasibility of ultra-high-resolution MicroCT imaging in elucidating important details of intracranial stent deployment as a basis for accurate CHD simulations and in enabling a structural and hemodynamic study of realistically deployed stents with different geometry and design.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Simulação por Computador / Tomografia Computadorizada por Raios X / Stents / Imageamento Tridimensional / Modelos Cardiovasculares Limite: Humans Idioma: En Revista: AJNR Am J Neuroradiol Ano de publicação: 2009 Tipo de documento: Article País de afiliação: Estados Unidos País de publicação: EEUU / ESTADOS UNIDOS / ESTADOS UNIDOS DA AMERICA / EUA / UNITED STATES / UNITED STATES OF AMERICA / US / USA

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Simulação por Computador / Tomografia Computadorizada por Raios X / Stents / Imageamento Tridimensional / Modelos Cardiovasculares Limite: Humans Idioma: En Revista: AJNR Am J Neuroradiol Ano de publicação: 2009 Tipo de documento: Article País de afiliação: Estados Unidos País de publicação: EEUU / ESTADOS UNIDOS / ESTADOS UNIDOS DA AMERICA / EUA / UNITED STATES / UNITED STATES OF AMERICA / US / USA