Cavotricuspid isthmus ablation with large-tip gold alloy versus platinum-iridium-tip electrode catheters.
Kardos, Attila; Foldesi, Csaba; Mihalcz, Attila; Szili-Torok, Tamas.
Afiliação
  • Kardos A; Gottsegen Gyorgy Hungarian Institute of Cardiology, Budapest, Hungary. dr.kardos@tonline.hu
Pacing Clin Electrophysiol ; 32 Suppl 1: S138-40, 2009 Mar.
Article em En | MEDLINE | ID: mdl-19250078
ABSTRACT

BACKGROUND:

Gold has excellent electrical conductive properties and creates deeper and wider lesions than platinum-iridium during radiofrequency (RF) ablation in vitro. We tested the maximum voltage-guided technique (MVGT) of cavotricuspid isthmus (CTI) ablation using two 8-mm tip catheters containing gold (group G) or platinum-iridium (group PI).

METHODS:

We enrolled 31 patients who underwent CTI ablation. In group G (n = 15) CTI ablation was performed with a gold-tip ablation catheter, while in group PI (n = 16) a platinum-iridium tip was used. Ablation was guided by CTI potentials with the highest amplitude until achievement of bidirectional block (BIB). If BIB was not achieved after 10 RF applications, RF was delivered via a 3.5-mm irrigated-tip catheter. Success rate, procedure duration, duration of fluoroscopic exposure, and number of RF applications were measured.

RESULTS:

BIB was achieved in all patients in group G, while in group PI an irrigated tip was used in four patients (0% vs 25%, P < 0.001). These four patients required a total of 21 additional RF applications (5.25 +/- 2.22). Procedure time (56.4 +/- 12 vs 73.1 +/- 15 minutes P < 0.05) and fluoroscopic explosure (4.9 +/- 2.3 vs 7.1 +/- 3.8 minutes, P < 0.01) were shorter in group G than in group PI. Mean number of RF applications was lower (4.6 +/- 1.9 vs 6.6 +/- 3.1 P < 0.001) and total RF duration shorter (280 +/- 117 vs 480 +/- 310 seconds) in group G than in group PI. No difference was observed in the number of recurrences at a 6 month-follow up (1 in group G vs 1 in group PI).

CONCLUSION:

Using the MVGT of CTI ablation, gold-tip catheters were associated with shorter procedural and fluoroscopic times, and fewer RF applications.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Flutter Atrial / Valva Tricúspide / Cateterismo / Ablação por Cateter / Eletrodos / Ouro Tipo de estudo: Clinical_trials Limite: Aged / Female / Humans / Male Idioma: En Revista: Pacing Clin Electrophysiol Ano de publicação: 2009 Tipo de documento: Article País de afiliação: Hungria

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Flutter Atrial / Valva Tricúspide / Cateterismo / Ablação por Cateter / Eletrodos / Ouro Tipo de estudo: Clinical_trials Limite: Aged / Female / Humans / Male Idioma: En Revista: Pacing Clin Electrophysiol Ano de publicação: 2009 Tipo de documento: Article País de afiliação: Hungria